Literature DB >> 11961640

Improved usability of a new handle design for laparoscopic dissection forceps.

M A Veelen1, D W Meijer, R H M Goossens, C J Snijders, J J Jakimowicz.   

Abstract

BACKGROUND: Recent studies have shown that the shape of most instrument handles causes user discomfort and that none of the handles currently available on the market satisfies all ergonomic criteria. Therefore, we designed a handle with improved usability; i.e., it is easier to manipulate and its use entails less risk of injury for the operator. The aim of this study was to demonstrate that the intended improvements of the new design were successful.
METHODS: Six handles representative of the types that are currently available were compared with the new version. Eight surgeons used the seven handles during a precision task and a rough task in a pelvi-trainer. A questionnaire and video analyses were used to assess the usability of the handles.
RESULTS: Based on the responses to the questionnaire, we determined that there were significant differences between the new handle and the Aesculap handle (means of rotation and angle between handle and shaft). No significant differences were found between the new handle and the five other handles. The video analyses (documenting extreme vs neutral wrist excursions) showed significant differences between the new handle and the six other handles.
CONCLUSION: Compared to the six existing handles, the new handle has significantly improved usability; in particular, its hinged design obviates the need to make extreme wrist excursions.

Mesh:

Year:  2001        PMID: 11961640     DOI: 10.1007/s004640090102

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  6 in total

1.  Improvement of the laparoscopic needle holder based on new ergonomic guidelines.

Authors:  M A van Veelen; D W Meijer; I Uijttewaal; R H M Goossens; C J Snijders; G Kazemier
Journal:  Surg Endosc       Date:  2003-03-06       Impact factor: 4.584

2.  Ergonomic problems encountered by the medical team related to products used for minimally invasive surgery.

Authors:  M A van Veelen; E A L Nederlof; R H M Goossens; C J Schot; J J Jakimowicz
Journal:  Surg Endosc       Date:  2003-05-06       Impact factor: 4.584

3.  Ergonomics of disposable handles for minimally invasive surgery.

Authors:  D Büchel; R Mårvik; B Hallabrin; U Matern
Journal:  Surg Endosc       Date:  2009-10-29       Impact factor: 4.584

4.  Initial experience using a robotic-driven laparoscopic needle holder with ergonomic handle: assessment of surgeons' task performance and ergonomics.

Authors:  Juan A Sánchez-Margallo; Francisco M Sánchez-Margallo
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-07-10       Impact factor: 2.924

5.  Rating and issues of mechanical anastomotic staplers in surgical practice: a survey of 241 Japanese gastroenterological surgeons.

Authors:  Emiko Kono; Yasuko Tomizawa; Tomoko Matsuo; Sachiyo Nomura
Journal:  Surg Today       Date:  2012-08-23       Impact factor: 2.549

6.  Ergonomic evaluation of a mechanical anastomotic stapler used by Japanese surgeons.

Authors:  Emiko Kono; Mitsunori Tada; Makiko Kouchi; Yui Endo; Yasuko Tomizawa; Tomoko Matsuo; Sachiyo Nomura
Journal:  Surg Today       Date:  2013-07-27       Impact factor: 2.549

  6 in total