Thin-section multidetector CT angiography of renal artery stents.

Jeffrey V Behar1, Rendon C Nelson, James P Zidar, David M DeLong, Tony P Smith.   

Abstract

OBJECTIVE: This study was undertaken as a pilot investigation to compare multidetector CT angiography with conventional catheter angiography for the visualization of the renal artery lumen after renal artery stent placement. SUBJECTS AND METHODS: CT angiography was performed within 24-48 hr of renal artery stent placement in 15 patients. Two patients had bilateral stents, resulting in a total of 17 stents. CT angiography was performed using a multidetector scanner and a bolus of IV contrast material with the scanning delay determined by a small-volume timing bolus. A volumetric data set was acquired through the stented arteries in the axial plane using a 4.0 x 1.25 mm detector configuration and a pitch of 3:1. The stent lumen diameter, as measured on direct CT angiography and curved multiplanar reformations in both the axial and coronal planes, was compared with that measured on catheter angiography.
RESULTS: The lumina of all 17 stents were well visualized and patent on both CT angiography and catheter angiography. Anatomic definition, including stent position and wall apposition in the renal artery, correlated well with catheter angiography. The diameter of the renal artery stent lumen measured on catheter angiography (mean, 5.9 +/- 1.3 mm) was greater than that on CT angiography (mean stent lumen diameter for direct axial plane was 4.6 +/- 1.0 mm, for curved multiplanar reformations in the axial plane was 4.3 +/- 1.0 mm, and for curved multiplanar reformations in the coronal plane was 4.4 +/- 1.0 mm) in 14 (82%) of 17 stents.
CONCLUSION: CT angiography produced interpretable multiplanar images of the renal artery, even with a metallic stent in place, and was adequate for determining stent patency. Compared with catheter angiography, the intrastent luminal diameter was underestimated in most patients who underwent CT angiography.

Entities:  

Mesh:

Year:  2002        PMID: 11959722     DOI: 10.2214/ajr.178.5.1781155

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  5 in total

1.  Computed Tomography Angiography of Carotid Stent. Comparison of Various Self-expandable Stent in a Phantom Model.

Authors:  C Sakai; N Sakai; T Okada; T Kuroiwa; H Ishihara; A Morizane; T Yano; H Kikuchi
Journal:  Interv Neuroradiol       Date:  2006-06-15       Impact factor: 1.610

Review 2.  Cross-sectional vascular imaging with CT and MR angiography.

Authors:  Hasan K Kabul; Klaus D Hagspiel
Journal:  J Nucl Cardiol       Date:  2006 May-Jun       Impact factor: 5.952

3.  Flash pulmonary oedema: accurate evaluation of the renal arteries with multislice computed tomography.

Authors:  G Morgan-Hughes; A J Marshall; C Roobottom
Journal:  Heart       Date:  2003-10       Impact factor: 5.994

4.  Congenital renal malrotation in ovarian cancer surgery: A case report.

Authors:  Lodovico Patrizi; Giacomo Corrado; Maria Saltari; Emilio Piccione; Enrico Vizza
Journal:  Gynecol Oncol Case Rep       Date:  2012-12-12

5.  Long-term follow-up after renal artery stenting.

Authors:  Robert A Bucek; Stefan Puchner; Markus Reiter; Albert Dirisamer; Erich Minar; Johannes Lammer
Journal:  Wien Klin Wochenschr       Date:  2003-11-28       Impact factor: 2.275

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.