Literature DB >> 11958911

Dosimetric characteristics of the MammoSite RTS, a new breast brachytherapy applicator.

Gregory K Edmundson1, Frank A Vicini, Peter Y Chen, Christina Mitchell, Alvaro A Martinez.   

Abstract

PURPOSE: A new device has been developed with the goal of making breast-conserving therapy more widely available, by making breast brachytherapy as monotherapy simpler and more accurate. This is the first published report on the dosimetric characteristics of this device.
MATERIALS AND METHODS: The experience of a single institution participating in a multi-institutional trial is reported. Twelve patients were enrolled, of whom 8 were treated. Computed tomography scans of all 12 patients were obtained before initiation of treatment. A commercial three-dimensional planning system was used for retrospective detailed analysis of dosimetry, including dose-volume histograms and, in selected patients, skin dose distributions. These data were compared to those of a similar group of patients, analyzed previously, that was treated with interstitial high-dose-rate brachytherapy.
RESULTS: Compared with interstitial brachytherapy, the MammoSite device resulted in a treatment dose that was less uniform, with a mean dose homogeneity index of 0.77 vs. 0.93. Coverage and reproducibility were improved, with a mean D90 (minimum dose to 90% of target volume) of 90.0% of prescribed dose (SD 0.5, range: 89.2-90.8%) vs. 69.8% (SD 7.3, range: 61.1-83.5%) for interstitial brachytherapy.
CONCLUSIONS: Reproducible placement was easily achieved in this very early experience, indicating a much smaller "learning curve" than for interstitial brachytherapy. Because skin doses can be increased with the use of this device, caution is indicated, although simple maneuvers can significantly reduce skin dose.

Entities:  

Mesh:

Year:  2002        PMID: 11958911     DOI: 10.1016/s0360-3016(01)02773-0

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  22 in total

1.  Post-lumpectomy intracavitary retention and lymph node targeting of (⁹⁹m)Tc-encapsulated liposomes in nude rats with breast cancer xenograft.

Authors:  Shihong Li; Beth Goins; William T Phillips; Marcela Saenz; Pamela M Otto; Ande Bao
Journal:  Breast Cancer Res Treat       Date:  2010-12-23       Impact factor: 4.872

Review 2.  Current role of modern radiotherapy techniques in the management of breast cancer.

Authors:  Gokhan Ozyigit; Melis Gultekin
Journal:  World J Clin Oncol       Date:  2014-08-10

Review 3.  Accelerated partial breast irradiation after conservative surgery for breast cancer.

Authors:  Henry M Kuerer; Thomas B Julian; Eric A Strom; H Kim Lyerly; Armando E Giuliano; Eleftherios P Mamounas; Frank A Vicini
Journal:  Ann Surg       Date:  2004-03       Impact factor: 12.969

Review 4.  Accelerated Partial Breast Irradiation (APBI): A review of available techniques.

Authors:  Christopher F Njeh; Mark W Saunders; Christian M Langton
Journal:  Radiat Oncol       Date:  2010-10-04       Impact factor: 3.481

5.  Immediate post-operative brachytherapy prior to irradiation and temozolomide for newly diagnosed glioblastoma.

Authors:  J Dawn Waters; Brent Rose; David D Gonda; Daniel J Scanderbeg; Michelle Russell; John F Alksne; Kevin Murphy; Bob S Carter; Joshua Lawson; Clark C Chen
Journal:  J Neurooncol       Date:  2013-05-15       Impact factor: 4.130

Review 6.  Current status and perspectives of brachytherapy for breast cancer.

Authors:  Csaba Polgár; Tibor Major
Journal:  Int J Clin Oncol       Date:  2009-02-20       Impact factor: 3.402

Review 7.  Accelerated partial breast irradiation for early-stage breast cancer: controversies and current indications for use.

Authors:  Lisa C Klepczyk; Kimberly S Keene; Jennifer F De Los Santos
Journal:  Curr Treat Options Oncol       Date:  2013-03

8.  On-Board Imager-based MammoSite treatment verification.

Authors:  Jadwiga Wojcicka; Rafael Yankelevich; Stephen Iorio; Alfred Tinger
Journal:  Med Biol Eng Comput       Date:  2007-07-12       Impact factor: 3.079

9.  Dose perturbation study in a multichannel breast brachytherapy device.

Authors:  Eric D Slessinger; Rodney Fletcher; Indra J Das
Journal:  J Contemp Brachytherapy       Date:  2011-12-30

10.  On the feasibility of treating to a 1.5 cm PTV with a commercial single-entry hybrid applicator in APBI breast brachytherapy.

Authors:  Kent A Gifford; Christopher L Nelson; Steven M Kirsner; Kelly D Kisling; Matthew T Ballo; Elizabeth S Bloom
Journal:  J Contemp Brachytherapy       Date:  2012-03-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.