Q Zhu1, N G Chen, D Piao, P Guo, X Ding.
Abstract
A total of 364 optical source-detector pairs were deployed uniformly over a 9 cm x 9 cm probe area initially, and then the total pairs were reduced gradually to 60 in experimental and simulation studies. For each source-detector configuration, three-dimensional (3-D) images of a 1-cm-diameter absorber of different contrasts were reconstructed from the measurements made with a frequency-domain system. The results have shown that more than 160 source-detector pairs are needed to reconstruct the absorption coefficient to within 60% of the true value and appropriate spatial and contrast resolution. However, the error in target depth estimated from 3-D images was more than 1 cm in all source-detector configurations. With the a priori target depth information provided by ultrasound, the accuracy of the reconstructed absorption coefficient was improved by 15% and 30% on average, and the beam width was improved by 24% and 41% on average for high- and low-contrast cases, respectively. The speed of reconstruction was improved by ten times on average.Year: 2001 PMID: 11958271 PMCID: PMC3568490 DOI: 10.1364/ao.40.003288
Source DB: PubMed Journal: Appl Opt ISSN: 1559-128X Impact factor: 1.980