Literature DB >> 11952444

Pulse contour analysis versus thermodilution in cardiac surgery patients.

H Rauch1, M Müller, F Fleischer, H Bauer, E Martin, B W Böttiger.   

Abstract

BACKGROUND: Previous studies have demonstrated that there is a lack of agreement between intermittent cold bolus thermodilution (ICO) and a semicontinuous method with dilution of heat (CCO) in cardiac surgical patients following hypothermic extracorporeal circulation (HCPB). Therefore, the aim of the present study was to compare both ICO and CCO with continuous pulse contour analysis (PCCO): a method based on a fundamentally different principle of determining cardiac output (CO).
METHODS: A prospective criterion standard study of 25 cardiac surgery patients undergoing HCPB. Cardiac output was determined using the three methods (ICO, CCO, and PCCO) before and after HCPB up to 12 h after arrival on the ICU. Bias and precision were evaluated.
RESULTS: A total of 380 triple determinations of CO could be analyzed. During the entire study period bias PCCO-ICO was -0.14 l*/min (precision 1.16 l*/min) and bias CCO-ICO was -0.40 l*/min (precision 1.25 l*/min). Up to 45 min after bypass PCCO agreed with ICO (bias -0.21 l*/min, precision 1.37 l*/min), while bias CCO-ICO was -1.30 l*/min (precision 1.45 l*/min).
CONCLUSION: The agreement between PCCO and ICO in contrast to CCO in the first 45 min after HCPB indicates that CCO underestimates CO during this period.

Entities:  

Mesh:

Year:  2002        PMID: 11952444     DOI: 10.1034/j.1399-6576.2002.460416.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  20 in total

1.  Reliability of continuous pulse contour cardiac output measurement during hemodynamic instability.

Authors:  Anders Johansson; Michelle Chew
Journal:  J Clin Monit Comput       Date:  2007-06-01       Impact factor: 2.502

Review 2.  [Intensive care management of critically ill adults with congenital heart disease].

Authors:  D Bettex; M Bosshart; P G Chassot; A Rudiger
Journal:  Med Klin Intensivmed Notfmed       Date:  2013-08-29       Impact factor: 0.840

Review 3.  [Haemodynamic monitoring in the perioperative phase. Available systems, practical application and clinical data].

Authors:  U Wittkowski; C Spies; M Sander; J Erb; A Feldheiser; C von Heymann
Journal:  Anaesthesist       Date:  2009-08       Impact factor: 1.041

4.  Pulse pressure: more than 100 years of changes in stroke volume.

Authors:  Maurizio Cecconi; Andrew Rhodes
Journal:  Intensive Care Med       Date:  2011-03-05       Impact factor: 17.440

Review 5.  Methods in pharmacology: measurement of cardiac output.

Authors:  Bart F Geerts; Leon P Aarts; Jos R Jansen
Journal:  Br J Clin Pharmacol       Date:  2011-03       Impact factor: 4.335

6.  [Accuracy of pulse contour cardiac index measurements during changes of preload and aortic impedance].

Authors:  T W Felbinger; M S Goepfert; T Goresch; A E Goetz; D A Reuter
Journal:  Anaesthesist       Date:  2005-08       Impact factor: 1.041

7.  [Non-invasive extended hemodynamic monitoring. Reduction of circulatory risk situations].

Authors:  M Bock; T Sturm; J Motsch
Journal:  Anaesthesist       Date:  2007-07       Impact factor: 1.041

8.  S3 guidelines for intensive care in cardiac surgery patients: hemodynamic monitoring and cardiocirculary system.

Authors:  M Carl; A Alms; J Braun; A Dongas; J Erb; A Goetz; M Goepfert; W Gogarten; J Grosse; A R Heller; M Heringlake; M Kastrup; A Kroener; S A Loer; G Marggraf; A Markewitz; D Reuter; D V Schmitt; U Schirmer; C Wiesenack; B Zwissler; C Spies
Journal:  Ger Med Sci       Date:  2010-06-15

9.  [Noninvasive determination of cardiac output in ventilated patients . Clinical evaluation of a simplified quick method].

Authors:  J Reutershan; T Kapp; K Unertl; R Fretschner
Journal:  Anaesthesist       Date:  2003-09       Impact factor: 1.041

10.  Stroke volume determination using transcardiopulmonary thermodilution and arterial pulse contour analysis in severe aortic valve disease.

Authors:  Martin Petzoldt; Carsten Riedel; Jan Braeunig; Sebastian Haas; Matthias S Goepfert; Hendrik Treede; Stephan Baldus; Alwin E Goetz; Daniel A Reuter
Journal:  Intensive Care Med       Date:  2013-01-04       Impact factor: 17.440

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.