Literature DB >> 11951811

Three-dimensional computerized stress analysis of commercially pure titanium and yttrium-partially stabilized zirconia implants.

Ralf-Joachim Kohal1, George Papavasiliou, Phophi Kamposiora, Aris Tripodakis, Jörg Rudolf Strub.   

Abstract

PURPOSE: The purpose of this study was to use three-dimensional finite element analysis to analyze stress distribution patterns in Re-Implant implants made of commercially pure titanium (cpTi) and yttrium-partially stabilized zirconia (YPSZ).
MATERIALS AND METHODS: Two three-dimensional finite element analysis models of a maxillary incisor with Re-Implant implants were made, surrounded by cortical and cancellous bone. A porcelain-fused-to-metal crown for the cpTi implant and a ceramic crown for the YPSZ implant were modeled. Stress levels were calculated according to the von Mises criteria.
RESULTS: Higher stresses were observed at the area where the implant entered the bone. Stresses were higher at the facial and lingual surfaces than the proximal ones. In cortical bone and at the junction of cortical and cancellous bone, stress distribution presented a pattern of alternating higher (4.0 to 5.0 MPa) and lower (1.3 to 2.0 MPa) stress areas. Higher stresses were found at the apical third of the implant-to-bone junction as well.
CONCLUSION: Re-Implant implants presented a pattern of low, well-distributed stresses along the entire implant-to-bone interface. YPSZ implants had very similar stress distribution to cpTi implants and may be viable esthetic alternatives, especially in maxillary anterior regions.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11951811

Source DB:  PubMed          Journal:  Int J Prosthodont        ISSN: 0893-2174            Impact factor:   1.681


  8 in total

1.  On the fracture of a zirconia ball head.

Authors:  C Piconi; G Maccauro; L Pilloni; W Burger; F Muratori; H G Richter
Journal:  J Mater Sci Mater Med       Date:  2006-03       Impact factor: 3.896

2.  Immediate, Non Submerged Root Analog Zirconia Implant in Single Rooted Tooth Replacement: Case Report with 2 years Follow Up.

Authors:  Amod Patankar; Rajesh Kshirsagar; Swapna Patankar; Sudhir Pawar
Journal:  J Maxillofac Oral Surg       Date:  2015-03-27

3.  Evaluation of Stress Generated with Different Abutment Materials and Angulations under Axial and Oblique Loading in the Anterior Maxilla: Three-Dimensional Finite Element Analysis.

Authors:  Sumedha Kapoor; Shobha Rodrigues; M Mahesh; Thilak Shetty; Umesh Pai; Sharon Saldanha; Puneeth Hedge; Satish Shenoy
Journal:  Int J Dent       Date:  2021-12-01

4.  Effects of Marginal Bone Loss Progression on Stress Distribution in Different Implant-Abutment Connections and Abutment Materials: A 3D Finite Element Analysis Study.

Authors:  Ching-Ping Lin; Yi-Ting Shyu; Yu-Ling Wu; Ming-Hsu Tsai; Hung-Shyong Chen; Aaron Yu-Jen Wu
Journal:  Materials (Basel)       Date:  2022-08-25       Impact factor: 3.748

Review 5.  Oral Tissue Interactions and Cellular Response to Zirconia Implant-Prosthetic Components: A Critical Review.

Authors:  Marcel F Kunrath; Saurabh Gupta; Felice Lorusso; Antonio Scarano; Sammy Noumbissi
Journal:  Materials (Basel)       Date:  2021-05-25       Impact factor: 3.623

6.  Multiple teeth replacement with endosseous one-piece yttrium-stabilized zirconia dental implants.

Authors:  Andrea-Enrico Borgonovo; Alberto Fabbri; Virna Vavassori; Rachele Censi; Carlo Maiorana
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2012-11-01

Review 7.  Zirconia in dental implantology: A review.

Authors:  Abhishek Apratim; Prashanti Eachempati; Kiran Kumar Krishnappa Salian; Vijendra Singh; Saurabh Chhabra; Sanket Shah
Journal:  J Int Soc Prev Community Dent       Date:  2015 May-Jun

8.  Biomechanical effects of original equipment manufacturer and aftermarket abutment screws in zirconia abutment on dental implant assembly.

Authors:  Yu-Ling Wu; Ming-Hsu Tsai; Hung-Shyong Chen; Yu-Tsen Chang; Tsai-Te Lin; Aaron Yu-Jen Wu
Journal:  Sci Rep       Date:  2020-10-27       Impact factor: 4.379

  8 in total