Quantification of errors in volume measurements of the caudate nucleus using magnetic resonance imaging.

Kilichan Gurleyik1, E Mark Haacke.   

Abstract

PURPOSE: To quantify the various sources of error in measuring the volume of the caudate nucleus and to understand these errors would lead to the standardization of the MRI protocol and would make the utility of data from around the world more viable in a global database.
MATERIALS AND METHODS: We collected data at four different sites all using a Siemens 1.5T Vision MR Scanner. In all cases the same 3D gradient-echo scans were used on a single volunteer and analyzed by a set of five observers.
RESULTS: The errors estimated were: system calibration (a random variation of up to 1.2%), partial volume error (a bias of up to 1.5% using isotropic resolution of 1 x 1 x 1 mm(3)), geometric distortion (a potential bias of 1%), intra-observer error (a random variation of up to 3%), effects of ringing (area biases of up to 7% when a zoom of 4 was used) and inter-observer error (with a bias of usually 5- 10% but sometimes as large as 16% among our five observers). Individual mean variations from one system to another differed by less than 5% (except for two observers at one site), consistent with a maximum error of 7% coming from the area bias due to limitations in the images themselves. We also measured the effect of variable resolution on the volume estimates and found that the measured volumes were consistent over a broad range of signal-to-noise-ratios (SNRs).
CONCLUSION: Given the observed dependence of the caudate volumes on SNR and resolution, if isotropic resolution is required because a complicated structure is being imaged, then the lower SNR suffered by collecting 1 x 1 x 1 mm(3) data at 1.5T still appears to be sufficient to make accurate volume measurements as long as the contrast-to-noise ratio (CNR) is on the order of 4:1. Based on our results, predictions are made as to what the best approach would be to improve the data acquisition scheme to keep individual errors under 2% and biases under 3.5%. We conclude that if users can be trained to identify the structure of interest in the same way, the inter-observer error could be reduced to that of intra-observer day-to-day error. Copyright 2002 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2002        PMID: 11948824     DOI: 10.1002/jmri.10083

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  5 in total

1.  In vivo basal ganglia volumetry through application of NURBS models to MR images.

Authors:  Giuseppe Anastasi; Giuseppina Cutroneo; Francesco Tomasello; Sebastiano Lucerna; AntonGiulio Vitetta; Placido Bramanti; Paolo Di Bella; Anna Parenti; Andrea Porzionato; Veronica Macchi; Raffaele De Caro
Journal:  Neuroradiology       Date:  2006-03-28       Impact factor: 2.804

2.  Measurement of Signal-to-Noise and Contrast-to-Noise in the fBIRN Multicenter Imaging Study.

Authors:  Vincent A Magnotta; Lee Friedman
Journal:  J Digit Imaging       Date:  2006-06       Impact factor: 4.056

3.  Reduced k-space acquisition to accelerate MR imaging of moving interventional instruments: a phantom study.

Authors:  Jens Christian Rump; Martin Jonczyk; Christian Jürgen Seebauer; Florian Streitparth; Felix Victor Güttler; Ulf Karl-Martin Teichgräber; Bernd Hamm
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-03-17       Impact factor: 2.924

4.  Determination of brain iron content in patients with Parkinson's disease using magnetic susceptibility imaging.

Authors:  Wei Zhang; Sheng-Gan Sun; Yu-Hong Jiang; Xian Qiao; Xin Sun; Yan Wu
Journal:  Neurosci Bull       Date:  2009-12       Impact factor: 5.203

5.  Displacement of brain regions in preterm infants with non-synostotic dolichocephaly investigated by MRI.

Authors:  Andrea U J Mewes; Lilla Zöllei; Petra S Hüppi; Heidelise Als; Gloria B McAnulty; Terrie E Inder; William M Wells; Simon K Warfield
Journal:  Neuroimage       Date:  2007-04-18       Impact factor: 6.556

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.