OBJECTIVE: To determine whether fibrin-coated central venous catheters have a higher infection rate, and spawn more septic emboli, than uncoated catheters after exposure to bacteremia.
DESIGN: Animal study comparing catheter infection and blood cultures of fibrin-coated and uncoated catheters exposed to bacteremia.
SETTING: Animal laboratory.
SUBJECTS: Adult male Sprague-Dawley rats.
INTERVENTIONS: A total of 210 rats had catheters placed with the proximal end buried subcutaneously. Rats were divided into three groups: tail vein bacterial injection on day 0 (no fibrin group) or on day 10 (fibrin group), or no injection/saline injection (control, n = 40). Bacterial injections were 1 x 108 colony forming units of either Staphylococcus epidermidis (n = 100) or Enterobacter cloacae (n = 60). Animals were killed 3 days after injection. Blood cultures were obtained via cardiac puncture, and catheters were removed via the chest. Half of the catheter was rolled onto agar and the other half was placed in trypticase soy broth. Plates and broth were incubated at 37 degrees C for 48 hrs. The presence of >15 colonies on roll plates, or growth in broth, was accepted as a positive sign of infection. Microscopy was performed on day 20-10 catheters. Thirty animals without catheters had bacterial injections and underwent blood culture 3 days after injection.
MEASUREMENTS AND MAIN RESULTS: Catheter infection with S. epidermidis occurred in 32% of roll plates and 80% of broth from the fibrin group vs. 4% and 20% from the no fibrin group (p <.01 for each). Catheter infection with E. cloacae occurred in 50% of roll plates and 80% of broth from the fibrin group vs. 0% and 12% from the no fibrin group (p <.01 for each). Positive blood cultures occurred in 47 of 68 animals from the fibrin group vs. 8 of 68 from the no fibrin group (p <.01). Microscopy showed a fibrin sheath on 20 of 20 catheters. Without catheters, 30 of 30 blood cultures were negative.
CONCLUSION: The fibrin sheath significantly enhanced catheter-related infection and persistent bacteremia.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11940768     DOI: 10.1097/00003246-200204000-00033

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  32 in total

1.  In Vitro and In Vivo Effectiveness of an Innovative Silver-Copper Nanoparticle Coating of Catheters To Prevent Methicillin-Resistant Staphylococcus aureus Infection.

Authors:  Myriam K S Ballo; Sami Rtimi; César Pulgarin; Nancy Hopf; Aurélie Berthet; John Kiwi; Philippe Moreillon; José M Entenza; Alain Bizzini
Journal:  Antimicrob Agents Chemother       Date:  2016-08-22       Impact factor: 5.191

2.  Surface modification of a polyethylene film for anticoagulant and anti-microbial catheter.

Authors:  Yingying Zheng; Jianjun Miao; Fuming Zhang; Chao Cai; Amanda Koh; Trevor J Simmons; Shaker A Mousa; Robert J Linhardt
Journal:  React Funct Polym       Date:  2016-03-01       Impact factor: 3.975

3.  Guidelines for the prevention of intravascular catheter-related infections.

Authors:  Naomi P O'Grady; Mary Alexander; Lillian A Burns; E Patchen Dellinger; Jeffrey Garland; Stephen O Heard; Pamela A Lipsett; Henry Masur; Leonard A Mermel; Michele L Pearson; Issam I Raad; Adrienne G Randolph; Mark E Rupp; Sanjay Saint
Journal:  Clin Infect Dis       Date:  2011-04-01       Impact factor: 9.079

Review 4.  A narrative review of long-term central venous access devices for the intensivist.

Authors:  David Moir; Andrew Bodenham
Journal:  J Intensive Care Soc       Date:  2017-12-12

Review 5.  Options and Limitations in Clinical Investigation of Bacterial Biofilms.

Authors:  Maria Magana; Christina Sereti; Anastasios Ioannidis; Courtney A Mitchell; Anthony R Ball; Emmanouil Magiorkinis; Stylianos Chatzipanagiotou; Michael R Hamblin; Maria Hadjifrangiskou; George P Tegos
Journal:  Clin Microbiol Rev       Date:  2018-04-04       Impact factor: 26.132

6.  Management of a Patient with Catheter-Related Bloodstream Infection.

Authors:  Charmaine E Lok
Journal:  Clin J Am Soc Nephrol       Date:  2017-07-05       Impact factor: 8.237

7.  Staphylococcus aureus coagulases are exploitable yet stable public goods in clinically relevant conditions.

Authors:  Urvish Trivedi; Jonas S Madsen; Jake Everett; Cody Fell; Jakob Russel; Jakob Haaber; Heidi A Crosby; Alexander R Horswill; Mette Burmølle; Kendra P Rumbaugh; Søren J Sørensen
Journal:  Proc Natl Acad Sci U S A       Date:  2018-11-21       Impact factor: 11.205

8.  Accumulation-associated protein enhances Staphylococcus epidermidis biofilm formation under dynamic conditions and is required for infection in a rat catheter model.

Authors:  Carolyn R Schaeffer; Keith M Woods; G Matt Longo; Megan R Kiedrowski; Alexandra E Paharik; Henning Büttner; Martin Christner; Robert J Boissy; Alexander R Horswill; Holger Rohde; Paul D Fey
Journal:  Infect Immun       Date:  2014-10-20       Impact factor: 3.441

9.  The role of tissue plasminogen activator use and systemic hypercoagulability in central line-associated bloodstream infections.

Authors:  Kinna Thakarar; Matthew Collins; Lana Kwong; Carol Sulis; Cathy Korn; Nahid Bhadelia
Journal:  Am J Infect Control       Date:  2014-02-20       Impact factor: 2.918

10.  Guidelines for the prevention of intravascular catheter-related infections: recommendations relevant to interventional radiology for venous catheter placement and maintenance.

Authors:  Donald L Miller; Naomi P O'Grady
Journal:  J Vasc Interv Radiol       Date:  2012-08       Impact factor: 3.464

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.