Literature DB >> 1192582

Factors influencing evaporation from sample cups and assessment of their effect on analytical error.

C A Burtis, J M Begovich, J S Watson.   

Abstract

We studied sample evaporation and its effect on analytical error. Several factors influencing evaporative loss have been identified and measured: environmental, instrumental, and operational factors, and the chemical and physical properties of the sample and its container. Such losses from several different types of sample cups have been measured, either chemically or gravimetrically, and compared with those calculated by using a model that allows evaporative loss from a cup of known geometry to be predicted under various environmental conditions. We discuss some steps that may be taken to minimize evaporative loss and give an example to demonstrate that analytical error from this source can be decreased to a routine 1--2% or less by selecting a particular cup design.

Mesh:

Substances:

Year:  1975        PMID: 1192582

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  2 in total

Review 1.  Misleading biochemical laboratory test results.

Authors:  A A Nanji
Journal:  Can Med Assoc J       Date:  1984-06-01       Impact factor: 8.262

2.  Evaluation of Serum Volume Losses During Long-Term Storage.

Authors:  Neal E Craft; Katherine S Epler; Therese A Butler; Willie E May; Regina G Ziegler
Journal:  J Res Natl Inst Stand Technol       Date:  1993 May-Jun
  2 in total