Literature DB >> 11921636

Automatic classification algorithms of the EEG monitor Narcotrend for routinely recorded EEG data from general anaesthesia: a validation study.

B Schultz1, U Grouven, A Schultz.   

Abstract

Impacts of hypnotic drugs on brain function are reflected in the EEG. The EEG monitor Narcotrend performs an automatic classification of the EEG using a scale which was proposed by Kugler for visual evaluation of the EEG. In this article the results of a validation study of the automatic classification algorithms implemented in the EEG monitor Narcotrend are presented. Visual and automatic classification of EEG data recorded in routine clinical practice were compared. The correlation between visual and automatic assessment was high (Spearman rank correlation r = 0.90, prediction probability Pk = 0.90) and a sufficient agreement between visual and automatic assessment was achieved for 92% of the analysed EEG epochs. The results of the study suggest that the automatic classification algorithms implemented in the EEG monitor Narcotrend yield a reliable assessment of the depth of hypnosis.

Entities:  

Mesh:

Year:  2002        PMID: 11921636     DOI: 10.1515/bmte.2002.47.1-2.9

Source DB:  PubMed          Journal:  Biomed Tech (Berl)        ISSN: 0013-5585            Impact factor:   1.411


  9 in total

Review 1.  [The Narcotrend monitor. Development and interpretation algorithms].

Authors:  B Schultz; S Kreuer; W Wilhelm; U Grouven; A Schultz
Journal:  Anaesthesist       Date:  2003-12       Impact factor: 1.041

2.  Correlation of Narcotrend Index, entropy measures, and spectral parameters with calculated propofol effect-site concentrations during induction of propofol-remifentanil anaesthesia.

Authors:  Ulrich Grouven; Frank A Beger; Barbara Schultz; Arthur Schultz
Journal:  J Clin Monit Comput       Date:  2004-08       Impact factor: 2.502

Review 3.  Using EEG to monitor anesthesia drug effects during surgery.

Authors:  Leslie C Jameson; Tod B Sloan
Journal:  J Clin Monit Comput       Date:  2006-12       Impact factor: 2.502

4.  Improving successful rate of transcranial electrical motor-evoked potentials monitoring during spinal surgery in young children.

Authors:  Junlin Yang; Zifang Huang; Haihua Shu; Yuguang Chen; Xinrui Sun; Weifeng Liu; Yunling Dou; Chaofan Xie; Xiang Lin; Yong Hu
Journal:  Eur Spine J       Date:  2011-08-30       Impact factor: 3.134

5.  [EEG-adjusted target-controlled infusion : Propofol target concentration with different doses of remifentanil].

Authors:  N Büttner; B Schultz; U Grouven; A Schultz
Journal:  Anaesthesist       Date:  2010-02       Impact factor: 1.041

6.  Comparison of Narcotrend Index, Bispectral Index, spectral and entropy parameters during induction of propofol-remifentanil anaesthesia.

Authors:  Arthur Schultz; Markus Siedenberg; Ulrich Grouven; Thomas Kneif; Barbara Schultz
Journal:  J Clin Monit Comput       Date:  2008-02-21       Impact factor: 2.502

7.  Peak and averaged bicoherence for different EEG patterns during general anaesthesia.

Authors:  Stacey Pritchett; Eugene Zilberg; Zheng Ming Xu; Paul Myles; Ian Brown; David Burton
Journal:  Biomed Eng Online       Date:  2010-11-20       Impact factor: 2.819

Review 8.  [Special aspects of analgosedation in cardiogenic shock patients].

Authors:  H Lemm; M Janusch; M Buerke
Journal:  Med Klin Intensivmed Notfmed       Date:  2016-01-25       Impact factor: 0.840

9.  Narcotrend-guided intraoperative care of a Trisomy 21 paediatric patient who underwent occipitocervical fusion.

Authors:  Evangeline Ko Villa; Dominic Villa; Rafael C Bundoc
Journal:  BMJ Case Rep       Date:  2020-02-11
  9 in total