| Literature DB >> 11918182 |
Satoshi Nishimura1, Toshifumi Yoshidome, Tomoyuki Tokuda, Masaru Mitsushio, Morihide Higo.
Abstract
A surface plasmon resonance (SPR) sensor was applied to analyses of some amine compounds (n-butylamine, isobutylamine, aniline, and N,N-dimethylaniline) by using a polymer film and an acid-base reaction in it. Poly(acrylamide) (PAA) was adopted as the polymer film and was immobilized on an Au film to prepare a sensor chip. Pivalic acid was entered into the PAA film as an acid. The PAA film with a thickness of 50 nm gave the highest sensitivity to the SPR sensor. Although water was better concerning the sensitivity for the SPR sensor as the solvent, ethanol was adopted because it dissolves well all of the amine compounds used. The Au film coated with the PAA film gave higher sensitivity for analyses of n-butylamine and isobutylamine, and lower sensitivity for analyses of aniline and N,N-dimethylaniline than an Au film without the PAA film. The PAA film containing pivalic acid gave 4-5 orders of magnitude higher sensitivity to the SPR sensor for analyses of all the amine compounds due to the reaction between pivalic acid and these amine compounds.Entities:
Year: 2002 PMID: 11918182 DOI: 10.2116/analsci.18.261
Source DB: PubMed Journal: Anal Sci ISSN: 0910-6340 Impact factor: 2.081