OBJECTIVE: This study evaluated the vacuum-suction needle (8- to 11-gauge) on an upright stereotactic machine with a lateral arm-support system. SUBJECTS AND METHODS: From July 1999 to August 2000, 185 core biopsies of the breast in 179 consecutive patients were planned in four institutions using 8- 11-gauge vacuum-suction probes on an upright stereotactic unit. Needles were stabilized and attached to the x, y, and z coordinates of the machines via a lateral arm-support system. Needle entry was made in the x-axis.
RESULTS: Five patients were canceled, and 180 biopsies were performed in 174 patients while the patients were in seated (n = 171) and lateral decubitus (n = 9) positions. An average of 9.5 cores were taken (range, 5-26 cores). Targeting was successful in 176 (98%) of 180 cores. Lesions were missed because of movement associated with Parkinson's disease (n = 1), or because the mass was obscured (n = 1) or calcifications were not in the core specimen (n = 2). Findings in 152 (84%) of the biopsies were benign and 28 (16%) were malignant. Forty-one lesions underwent surgical excision and 106 underwent mammographic follow-up. Discordance was 4% (6/147). Complications included vasovagal reactions (n = 10, 5.6%), bleeding (n = 5, 3%), hematomas (n = 3, 1.7%), vomiting (n = 1, 0.6%), and technical failure (n = 1, 0.6%).
CONCLUSION: Vacuum-suction needle core biopsies can be successfully performed on an upright stereotactic machine with a lateral arm attachment. Thinly compressed breasts and lesions located near the chest wall are well sampled. The vasovagal rate is higher than that on a prone table but is acceptable.

Entities:  

Mesh:

Year:  2002        PMID: 11906893     DOI: 10.2214/ajr.178.4.1781017

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  3 in total

1.  Stereotactic biopsy of the breast using a decubitus table: comparison of histologic underestimation rates between 11- and 8-gauge vacuum-assisted breast biopsy.

Authors:  Kyoung Eun Lee; Hak Hee Kim; Hee Jung Shin; Joo Hee Cha
Journal:  Springerplus       Date:  2013-10-22

Review 2.  Methodological aspects of 99mTc-sestamibi guided biopsy in breast cancer.

Authors:  A Collarino; R A Valdés Olmos; A F van der Hoeven; L M Pereira Arias-Bouda
Journal:  Clin Transl Imaging       Date:  2016-07-16

3.  Does lateral arm technique decrease the rate of clip migration in stereotactic and tomosynthesis-guided biopsies?

Authors:  Olena Weaver; Ethan O Cohen; Rachel E Perry; Hilda H Tso; Kanchan Phalak; Ashmitha Srinivasan; Roland Bassett; Jessica W T Leung
Journal:  Insights Imaging       Date:  2021-12-20
  3 in total