Accuracy of the avosure PT pro system compared with a hospital laboratory standard.

Janene M Rigelsky1, Hae Mi Choe, Dawn M Curtis, Marcia J Brosnan, Sonya Mitrovich, Daniel S Streetman.   

Abstract

OBJECTIVE: To compare international normalized ratio (INR) values obtained using the AvoSure PT Pro point-of-care (POC) system with those obtained using a standard laboratory method.
METHODS: Forty-one INR values obtained from the POC system were compared with those obtained from a standard laboratory method. The POC method was evaluated for both laboratory and clinical agreement. To evaluate laboratory agreement, various analyses were used, including mean-squared prediction error (MSE) and mean prediction error (ME), Bland-Altman analysis, correlation, and paired t-test comparing group INR means. For clinical accuracy, discrepant pairs were identified and evaluated to determine whether dosage adjustments would have been needed based on values obtained.
RESULTS: The POC system demonstrated modest precision (MSE = 0.147, 95% CI 0.065 to 0.228) and relatively little bias (ME = 0.090, 95% CI -0.025 to 0.205). Bland-Altman analysis also suggested good agreement at average INRs from 2.0 to 3.0. At average INR values >3.0, the POC system consistently overestimated INR. Values obtained with the POC system were significantly correlated with those obtained from the hospital laboratory (r = 0.77; p < 0.001). Similarly, mean +/- SD POC INR did not differ significantly from the laboratory-determined INR (2.45+/-0.59 vs. 2.37+/-0.48, respectively; p = 0.176). Regarding clinical accuracy, the values clinically agreed in 85.4% of the cases.
CONCLUSIONS: The AvoSure PT Pro POC system appears to be useful for INR values within the 2.0-3.0 range, but values outside of this range should probably be confirmed with a standard laboratory method.

Mesh:

Substances:

Year:  2002        PMID: 11895047     DOI: 10.1345/aph.1A253

Source DB:  PubMed          Journal:  Ann Pharmacother        ISSN: 1060-0280            Impact factor:   3.154


  2 in total

1.  Administration of tissue plasminogen activator without coagulation results in a Chinese population.

Authors:  Xiaoming Qin; Songyao Zhao; Liujie Yin; Hailing Dou; Jie Chen; Yifan Wang; Mingzhe Li; Ruifang Chen; Jing Fu; Wei Liu; Xin Liu; Gaiqing Yang; Runqing Wang; Xinzhou Jia; Shufang Bu; Dongpu Ma; Baoyu Wang; Shize Li
Journal:  Neurol Sci       Date:  2018-01-03       Impact factor: 3.307

2.  Evaluation of patient perceptions and outcomes related to anticoagulation point-of-care testing in ambulatory care clinics.

Authors:  Amy N Thompson; Kelly R Ragucci; Joli D Fermo; Heather P Whitley
Journal:  Pharm Pract (Granada)       Date:  2009-03-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.