[Clinical utilization of the coronary pressure wire].

Ramón López-Palop1, Eduardo Pinar, Iñigo Lozano, Pilar Carrillo, Rocío Cortés, Francisco Picó, Mariano Valdés.   

Abstract

INTRODUCTION: Earlier studies have established the value of coronary pressure wires for diagnosing and monitoring the treatment of patients with coronary artery disease. In this study we demonstrated their usefulness in the daily clinical practice of a catheterization laboratory.
MATERIAL AND METHODS: A retrospective study of the use of pressure wires in our laboratory between October 1998 and November 2000. The pressure wire was inserted whenever the interventional cardiologist considered it to be indicated. In all cases, pressures were recorded with a Waveguide Cardiometrics 0.014 guide (Endosonics) and hyperemia was induced by intracoronary adenosine.
RESULTS: Two hundred fifty-three lesions were studied in 190 patients. Indications were functional evaluation of lesions of intermediate severity for 82% (9% intrastent restenoses); guidance of balloon PTCA for 5%; and fulfillment of a research protocol for 13%. Twenty-six percent of lesions considered to be of moderate severity based on angiography were treated as a consequence of the pressures measured by the wire. A decision to begin or continue a procedure was based on wire pressures in 24% and intervention was avoided in 60%. No major complications attributable to the wire were observed. A lesion was dissected in one patient (0.5%) but it was treated without consequences. Twenty pressure wires (11%) failed to work properly during the procedure, fourteen of them (7%) before insertion. The wire could not be advanced across the lesion in one case.
CONCLUSIONS: The pressure wire is useful in the daily clinical practice of a catheterization laboratory. Its most common indication is the evaluation of lesions of intermediate or unknown severity, and use is associated with few complications.

Entities:  

Mesh:

Year:  2002        PMID: 11893316     DOI: 10.1016/s0300-8932(02)76593-7

Source DB:  PubMed          Journal:  Rev Esp Cardiol        ISSN: 0300-8932            Impact factor:   4.753


  3 in total

1.  Adequate intracoronary adenosine doses to achieve maximum hyperaemia in coronary functional studies by pressure derived fractional flow reserve: a dose response study.

Authors:  R Lopez-Palop; D Saura; E Pinar; I Lozano; F Pérez-Lorente; F Picó; M Valdez
Journal:  Heart       Date:  2004-01       Impact factor: 5.994

2.  Intravascular pressure measurements in feeding pedicles of brain arteriovenous malformations.

Authors:  H Henkes; T F Gotwald; S Brew; E Miloslavski; F Kämmerer; D Kühne
Journal:  Neuroradiology       Date:  2005-11-22       Impact factor: 2.804

3.  Evaluating the impact of fractional flow reserve-guided percutaneous coronary intervention in intermediate coronary artery lesions on the mode of treatment and their outcomes: An Iranian experience.

Authors:  Alireza Khosravi; Mohammad Reza Pourbehi; Masoud Pourmoghaddas; Afshin Ostovar; Mohammad Reza Akhbari; Fereshteh Ziaee-Bideh; Jafar Golshahi; Shahin Shirani
Journal:  ARYA Atheroscler       Date:  2015-03
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.