Positive pressure ventilation during fibreoptic intubation: comparison of the laryngeal mask airway, intubating laryngeal mask and endoscopy mask techniques.

K Aoyama1, E Yasunaga, I Takenaka, T Kadoya, T Sata, A Shigematsu.   

Abstract

BACKGROUND: The efficacy of delivery of mechanical ventilation through different airway devices during fibreoptic intubation is not known.
METHODS: We compared the laryngeal mask airway (LMA), intubating laryngeal mask (ILM) and endoscopy mask for positive pressure ventilation (PPV) during fibreoptic intubation. In 80 adult paralysed patients, fibreoptic intubation was performed during PPV using a combination of a size 3 or 4 LMA with a 6.0 mm nasal RAE tracheal tube (LMA3/4 group; n=22), a size 5 LMA with a 7.0 mm nasal RAE tube (LMA5 group; n=18), an ILM with an 8.0 mm special reinforced tracheal tube (ILM group; n=20) or an endoscopy mask (Patil mask) with a 7.5 mm standard tracheal tube (Patil group; n=20). The inspiratory and expiratory tidal volumes (VI and VE) with a ventilation pressure of 20 cm H2O were measured using a pneumotachograph.
RESULTS: Mean VE values during fibreoptic intubation in the LMA5 [5.3 (SD 1.5) ml kg(-1)] and ILM [7.1 (2.3) ml kg(-1)] groups were greater than in the LMA3/4 group [2.6 (1.0) ml kg(-1), P<0.0001]. The mean VE was greater in the Patil group [20.6 (4.9) ml kg(-1)] than in the other three groups (P<0.0001). Gastric insufflation during intubation was more frequent in the Patil group (30%) than in the other three groups (4.5-5.6%) (P<0.05).
CONCLUSION: PPV is possible with the LMA, ILM or endoscopy mask during fibreoptic intubation. With an airway pressure of 20 cm H2O, ventilation during intubation using a size 3 or 4 LMA was almost insufficient, while ventilation using a size 5 LMA or an ILM was almost acceptable. Ventilation during intubation with the endoscopy mask was greater than that with the LMA or ILM, but gastric insufflation was more frequent.

Entities:  

Mesh:

Year:  2002        PMID: 11878657     DOI: 10.1093/bja/88.2.246

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  3 in total

1.  JSA airway management guideline 2014: to improve the safety of induction of anesthesia.

Authors: 
Journal:  J Anesth       Date:  2014-07-03       Impact factor: 2.078

2.  Intubating condition, hemodynamic parameters and upper airway morbidity: A comparison of intubating laryngeal mask airway with standard direct laryngoscopy.

Authors:  J Kavitha; Debendra Kumar Tripathy; Sandeep Kumar Mishra; Gayatri Mishra; L J Chandrasekhar; P Ezhilarasu
Journal:  Anesth Essays Res       Date:  2011 Jan-Jun

3.  Tracheal Intubation with Aura-i and aScope-2: How to Minimize Apnea Time in an Unpredicted Difficult Airway.

Authors:  Vittorio Pavoni; Valentina Froio; Alessandra Nella; Martina Simonelli; Lara Gianesello; Andrew Horton; Luca Malino; Massimo Micaglio
Journal:  Case Rep Anesthesiol       Date:  2015-01-06
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.