Abstract
A novel triiodide ion-selective electrode based on a clotrimazole-triiodide ion pair as a membrane carrier was prepared. It has a linear response to triiodide from 8 x 10(-6) to 5 x 10(-3) M with a slope of -68.9 mV per decade and a detection limit of 5 x 10(-6) M. The electrode response is independent of the pH of the solution in the pH range 2-9. It has a very short response time and can be used for at least 3 months without any considerable divergence in the potentials. The proposed sensor revealed very good selectivities for I3- over a variety of other anions. It was used as an indicator electrode in the potentiometric titration of triiodide ions and in an indirect potentiometric determination of clotrimazole in pharmaceutical preparations.Entities:
Mesh:
Substances:
Year: 2002 PMID: 11874113 DOI: 10.2116/analsci.18.133
Source DB: PubMed Journal: Anal Sci ISSN: 0910-6340 Impact factor: 2.081