Literature DB >> 11870847

Multifunctional interventional devices for MRI: a combined electrophysiology/MRI catheter.

Robert C Susil1, Christopher J Yeung, Henry R Halperin, Albert C Lardo, Ergin Atalar.   

Abstract

The design and application of a two-wire electrophysiology (EP) catheter that simultaneously records the intracardiac electrogram and receives the MR signal for active catheter tracking is described. The catheter acts as a long loop receiver, allowing for visualization of the entire catheter length while simultaneously behaving as a traditional two-wire EP catheter, allowing for intracardiac electrogram recording and ablation. The application of the device is demonstrated by simultaneously tracking the catheter and recording the intracardiac electrogram in canine models using 7 and 10 frame/sec real-time imaging sequences. Using solely MR imaging, the entire catheter was visualized and guided from the jugular vein into the cardiac chambers, where the intracardiac electrogram was recorded. By combining several functions in a single, simple structure, the excellent tissue contrast and functional imaging capabilities of MR can be used to improve the efficacy of EP interventions. This catheter will facilitate MR-guided interventions and demonstrates the design of multifunctional interventional devices for use in MRI. Copyright 2002 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2002        PMID: 11870847     DOI: 10.1002/mrm.10088

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  22 in total

1.  Real-time accelerated interactive MRI with adaptive TSENSE and UNFOLD.

Authors:  Michael A Guttman; Peter Kellman; Alexander J Dick; Robert J Lederman; Elliot R McVeigh
Journal:  Magn Reson Med       Date:  2003-08       Impact factor: 4.668

Review 2.  Cardiovascular interventional magnetic resonance imaging.

Authors:  Robert J Lederman
Journal:  Circulation       Date:  2005-11-08       Impact factor: 29.690

3.  Real-time magnetic resonance imaging to guide pediatric endovascular procedures.

Authors:  A N Raval; R J Lederman
Journal:  Pediatr Cardiol       Date:  2005 May-Jun       Impact factor: 1.655

Review 4.  Real-time, Interactive MRI for cardiovascular interventions.

Authors:  Elliot R McVeigh; Michael A Guttman; Peter Kellman; Amish N Raval; Robert J Lederman
Journal:  Acad Radiol       Date:  2005-09       Impact factor: 3.173

Review 5.  Magnetic resonance imaging and radiofrequency ablations.

Authors:  T Dickfeld
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2007-09

Review 6.  Interventional cardiovascular magnetic resonance imaging.

Authors:  Venkatesh K Raman; Robert J Lederman
Journal:  Trends Cardiovasc Med       Date:  2007-08       Impact factor: 6.677

7.  Comparison of three artificial models of the magnetohydrodynamic effect on the electrocardiogram.

Authors:  Julien Oster; Raul Llinares; Stephen Payne; Zion Tsz Ho Tse; Ehud Jeruham Schmidt; Gari D Clifford
Journal:  Comput Methods Biomech Biomed Engin       Date:  2014-04-24       Impact factor: 1.763

8.  Voltage-based device tracking in a 1.5 Tesla MRI during imaging: initial validation in swine models.

Authors:  Ehud J Schmidt; Zion T H Tse; Tobias R Reichlin; Gregory F Michaud; Ronald D Watkins; Kim Butts-Pauly; Raymond Y Kwong; William Stevenson; Jeffrey Schweitzer; Israel Byrd; Charles L Dumoulin
Journal:  Magn Reson Med       Date:  2014-03       Impact factor: 4.668

Review 9.  Interventional cardiovascular magnetic resonance imaging: a new opportunity for image-guided interventions.

Authors:  Christina E Saikus; Robert J Lederman
Journal:  JACC Cardiovasc Imaging       Date:  2009-11

10.  Interactive real-time mapping and catheter ablation of the cavotricuspid isthmus guided by magnetic resonance imaging in a porcine model.

Authors:  Boris A Hoffmann; Andreas Koops; Thomas Rostock; Kai Müllerleile; Daniel Steven; Roman Karst; Mark U Steinke; Imke Drewitz; Gunnar Lund; Susan Koops; Gerhard Adam; Stephan Willems
Journal:  Eur Heart J       Date:  2009-11-05       Impact factor: 29.983

View more