Catheter-tip fixation of a percutaneously implanted port-catheter system to prevent dislocation.

T Yamagami1, T Nakamura, T Yamazaki, S Iida, T Kato, T Nishimura.   

Abstract

The aim of this study was to evaluate the incidence of catheter tip dislocation in patients with percutaneously implanted port-catheters for hepatic arterial chemotherapy with catheter tip fixation. Forty-seven patients (31 men and 16 women; mean age 66 years) with unresectable advanced liver cancers (primary liver cancer, n=19; metastatic liver cancer, n=28) underwent percutaneously implantable port-catheter system placement with the tip fixed at the gastroduodenal artery with coils and side hole opened at the common hepatic artery. In 39 patients, n-butyl cyanoacrylate (NBCA) mixed with Lipiodol was added for fixation. The position of the side hole after the indwelling port-catheter system was investigated, and the correction method in cases with catheter dislocation was determined. In 2 (25%) of the 8 patients without NBCA fixation, dislocation of the catheter was noted, in contrast to none (0%) of 37 patients with NBCA fixation. Two patients in whom NBCA was used could not undergo long-term intra-arterial chemotherapy because of hepatic arterial thrombotic occlusion which occurred after placement of the indwelling catheter, and were excluded from the evaluation. Fixation of the catheter tip with combined use of coils and NBCA--Lipiodol mixture to the gastroduodenal artery is important to prevent dislocation of the port-catheter system.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11870447     DOI: 10.1007/s003300100943

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  7 in total

1.  A simplified technique of percutaneous hepatic artery port-catheter insertion for the treatment of advanced hepatocellular carcinoma with portal vein invasion.

Authors:  Sun Young Choi; Ah Hyun Kim; Kyung Ah Kim; Jong Yun Won; Do Yun Lee; Kwang-Hun Lee
Journal:  Korean J Radiol       Date:  2010-10-29       Impact factor: 3.500

2.  Can microcatheters be cleaned for reuse after NBCA embolization? Cleaning technique with gelatin sponge particles.

Authors:  Takuji Araki; Hiroki Okada; Hiroshi Onishi
Journal:  Jpn J Radiol       Date:  2015-06-30       Impact factor: 2.374

3.  Dysfunction of a port-catheter system percutaneously implanted for hepatic arterial infusion chemotherapy due to a fibrin sheath.

Authors:  Takuji Yamagami; Tomohiro Matsumoto; Rika Yoshimatsu; Tsunehiko Nishimura
Journal:  Clin J Gastroenterol       Date:  2008-12-25

4.  Fixation methods for implantable port chamber: comparative study using glue, self-stabilizing leg and suture fixations in rabbits.

Authors:  Hyoung Il Na; Hyung Jin Shim; Byung Kook Kwak; Hyeon Joo Kim; Yong Cheol Lee
Journal:  Korean J Radiol       Date:  2004 Oct-Dec       Impact factor: 3.500

5.  J-tipped guidewire as a target for puncture of the subclavian artery in the placement of a reservoir port and catheter system.

Authors:  Yukihiro Hama; Kohzoh Makita; Shoichi Kusano
Journal:  Eur Radiol       Date:  2003-11-06       Impact factor: 5.315

6.  Value of Micronester coils in port-catheter implantation for continuous hepatic arterial infusion chemotherapy with fixed catheter tip method.

Authors:  Takuji Yamagami; Takeharu Kato; Tatsuya Hirota; Rika Yoshimatsu; Tomohiro Matsumoto; Robert I White; Tsunehiko Nishimura
Journal:  Eur Radiol       Date:  2007-07-10       Impact factor: 5.315

7.  Guidelines for the use of NBCA in vascular embolization devised by the Committee of Practice Guidelines of the Japanese Society of Interventional Radiology (CGJSIR), 2012 edition.

Authors:  Yoshito Takeuchi; Hiroyuki Morishita; Yozo Sato; Shingo Hamaguchi; Noriaki Sakamoto; Hiroyuki Tokue; Takafumi Yonemitsu; Kenji Murakami; Hiroyasu Fujiwara; Keitaro Sofue; Toshi Abe; Hideyuki Higashihara; Yasuo Nakajima; Morio Sato
Journal:  Jpn J Radiol       Date:  2014-06-03       Impact factor: 2.374

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.