A calibration study of therapeutic ultrasound units.

Paul A Artho1, Jason G Thyne, Barry P Warring, Chris D Willis, Jean-Michel Brismée, Neal S Latman.   

Abstract

BACKGROUND AND
PURPOSE: Physiological effects of therapeutic ultrasound (US) are dependent on the intensity and duration of application. The purpose of this study was to test US machines used in clinical settings for proper calibration of time and power output.
METHODS: Measurements of power output and timer accuracy were obtained from 83 US units in clinical use. The machines were tested at 4 intensity settings (0.5, 1.0, 1.5, and 2.0 W/cm2) using a continuous waveform and a 1-MHz frequency. The measured intensities were converted to percentages of error and compared with the +/-20% standard.
RESULTS: Of the machines tested, 32 (39%) were outside the calibration standard for at least one output setting. Of these machines, 15 (18%) were above the +20% standard, and 17 (21%) were below the -20% standard for at least one output setting. Of the 32 machines outside the standard, 26 (31%) were outside the standard for 2 or more settings, and 3 (4%) produced no output at any of the settings. Of the mechanical timers tested, 7 (28%) were outside of the +/-10% standard for timer accuracy at the 5-minute interval, and 6 (24%) were outside of the standard at the 10-minute interval. All digital timers tested were within the standard. DISCUSSION AND
CONCLUSION: More than one third of machines tested in this study were outside the standard for power output, and approximately one fourth of the mechanical timers were outside the standard. Therefore, further improvements in the accuracy of US machine calibration are needed.

Mesh:

Year:  2002        PMID: 11869154

Source DB:  PubMed          Journal:  Phys Ther        ISSN: 0031-9023


  8 in total

1.  ELECTROPHYSICAL AGENTS - Contraindications And Precautions: An Evidence-Based Approach To Clinical Decision Making In Physical Therapy.

Authors: 
Journal:  Physiother Can       Date:  2011-01-05       Impact factor: 1.037

2.  Frequency-induced morphology alterations in microconfined biological cells.

Authors:  Hritwick Banerjee; Bibhas Roy; Kaustav Chaudhury; Babji Srinivasan; Suman Chakraborty; Hongliang Ren
Journal:  Med Biol Eng Comput       Date:  2018-11-10       Impact factor: 2.602

Review 3.  A new direction for ultrasound therapy in sports medicine.

Authors:  Stuart J Warden
Journal:  Sports Med       Date:  2003       Impact factor: 11.136

4.  Variability in effective radiating area and output power of new ultrasound transducers at 3 MHz.

Authors:  Lennart D Johns; Stephen J Straub; Samuel M Howard
Journal:  J Athl Train       Date:  2007 Jan-Mar       Impact factor: 2.860

5.  Design of a Thermoacoustic Sensor for Low Intensity Ultrasound Measurements Based on an Artificial Neural Network.

Authors:  Jida Xing; Jie Chen
Journal:  Sensors (Basel)       Date:  2015-06-23       Impact factor: 3.576

6.  Assessing heating distribution by therapeutic ultrasound on bone phantoms and in vitro human samples using infrared thermography.

Authors:  Gabriella Sellani; Dalila Fernandes; Abigail Nahari; Melissa Fabrício de Oliveira; Christiana Valois; Wagner C A Pereira; Christiano B Machado
Journal:  J Ther Ultrasound       Date:  2016-04-05

7.  Laser light sources for photobiomodulation: The role of power and beam characterization in treatment accuracy and reliability.

Authors:  Carlos Eduardo Girasol; Guilherme de Araújo Braz; Luciano Bachmann; Jonathan Celli; Rinaldo Roberto de Jesus Guirro
Journal:  PLoS One       Date:  2022-03-30       Impact factor: 3.240

8.  A randomised controlled trial of 'clockwise' ultrasound for low back pain.

Authors:  Adriaan Louw; Kory Zimney; Merrill R Landers; Mark Luttrell; Bob Clair; Joshua Mills
Journal:  S Afr J Physiother       Date:  2016-07-28
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.