Literature DB >> 11861345

Active warming of saline or blood is ineffective when standard infusion tubing is used: an experimental study.

Bruno Bissonnette1, Olivier Paut.   

Abstract

PURPOSE: To determine the effect of infusion rate, tubing length and fluid composition on the temperature of the infusate reaching the distal end of an infusion tubing with and without active fluid warming.
METHODS: Warmed normal saline (W-NS) and packed red blood cells (W-PRBC), were infused with a fluid warmer through a modified infusion set. The fluids were delivered at eight infusion rates from 50 to 999 mL x hr(-1). The infusate temperature was monitored at 20 cm intervals on the iv tubing. The same temperature monitoring protocol was applied to PRBC without warmer (NoW-PRBC).
RESULTS: In W-NS and W-PRBC groups, there was a decrease in the infusate temperature, at each flow rate, from the drip chamber to the distal end of tubing ( P <0.001). In NoW-PRBC group, there was a rapid increase in the infusate temperature from the bag to the drip chamber ( P <0.001). Thereafter, there was no change in temperature, except at the 999 mL x hr(-1) infusion rate, where a slight increase in the infusate temperature throughout the tubing was shown. In W-NS and W-PRBC groups increasing the flow rate produced a significant increase in the infusate temperature, at each measurement point ( P <0.001). In the NoW-PRBC group, increasing the flow rate did not alter the infusate temperature. The fluid composition did not influence the infusate temperature.
CONCLUSION: There is an important heat exchange within the tubing, which is aggravated at low flow rates. At infusion rates appropriate for pediatric anesthesia the clinical and economic value of fluid warming without the use of heated extension tubing is questionable.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11861345     DOI: 10.1007/BF03020526

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  5 in total

1.  [AUTOLINE. Effectiveness of an infusion warmer concept].

Authors:  J Schnoor; I Weber; S Macko; R Rossaint
Journal:  Anaesthesist       Date:  2005-09       Impact factor: 1.041

2.  Experimental comparison of performances of Mega Acer Kit, Ranger and ThermoSens according to flow rates and distances.

Authors:  Hong Ju Seo; Sang Hun Kim; Tae Hun An; Dong Joon Kim
Journal:  J Clin Monit Comput       Date:  2017-02-07       Impact factor: 2.502

3.  Clinical evaluation of a newly designed fluid warming kit on fluid warming and hypothermia during spinal surgery.

Authors:  Ki Tae Jung; Sang Hun Kim; Keum Young So; Hyeong Jin So; Soo Bin Shim
Journal:  Korean J Anesthesiol       Date:  2015-09-30

4.  Heating capabilities of small fluid warming systems.

Authors:  Norbert Zoremba; Christian Bruells; Rolf Rossaint; Thomas Breuer
Journal:  BMC Anesthesiol       Date:  2018-07-28       Impact factor: 2.217

5.  Beneficial effect of fluid warming in elderly patients with bladder cancer undergoing Da Vinci robotic-assisted laparoscopic radical cystectomy.

Authors:  Jianwei Luo; Lin Zhou; Shaoman Lin; Wenchan Yan; Lijuan Huang; Sihua Liang
Journal:  Clinics (Sao Paulo)       Date:  2020-04-17       Impact factor: 2.365