Removal torques of conical, tapered implant abutments: the effects of anodization and reduction of surface area.

Rachel S Squier1, Walter J Psoter, Thomas D Taylor.   

Abstract

PURPOSE: To examine the effects of anodization (surface coating) and reduction of internal Morse taper surface area on the reversal torque values of Straumann ITI dental implants and abutments.
MATERIALS AND METHODS: Eighty ITI solid screw implants were mated with corresponding 5.5-mm solid abutments. The assemblies were divided into 4 test groups of 20 specimens. All abutments were torque tightened into the implant to 35 Ncm. Half of the abutments were anodized and half were in their as-machined state. Each of these 2 groups included half of the implants with the standard internal Morse taper configuration and half with the synOcta (Straumann USA, Waltham, MA) internal positioning interface (indexed). Torque removal testing was then performed on the assemblies. The 4 groups were compared statistically to examine the effect of the 2 variables (anodization and reduction in surface area). Scheffe's test for multiple comparisons was used to compare groups at an adjusted significance level of < or = .05.
RESULTS: Torque removal of all specimens revealed that the indexed implant with the non-anodized abutment demonstrated superior removal torque. DISCUSSION: The indexed and standard implants with anodized abutments, and the standard implant with the non-anodized abutment had lower reversal torque values.
CONCLUSION: The addition of the indexed internal surface to the ITI implant did not have deleterious effect on the resistance to loosening of standard solid abutments.

Mesh:

Substances:

Year:  2002        PMID: 11858571

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  9 in total

1.  A comparative biomechanical study of original and compatible titanium bases: evaluation of screw loosening and 3D-crown displacement following cyclic loading analysis.

Authors:  Rimantas Ožiūnas; Jurgina Sakalauskienė; Darius Jegelevičius; Gintaras Janužis
Journal:  J Adv Prosthodont       Date:  2022-04-27       Impact factor: 1.989

2.  The change of rotational freedom following different insertion torques in three implant systems with implant driver.

Authors:  Joo-Hyun Kwon; Chong-Hyun Han; Sun-Jai Kim; Jae-Seung Chang
Journal:  J Adv Prosthodont       Date:  2009-03-31       Impact factor: 1.904

3.  Complication incidence of two implant systems up to six years: a comparison between internal and external connection implants.

Authors:  Sung-Wook Chae; Young-Sung Kim; Yong-Moo Lee; Won-Kyung Kim; Young-Kyoo Lee; Su-Hwan Kim
Journal:  J Periodontal Implant Sci       Date:  2015-02-25       Impact factor: 2.614

4.  Stability of external and internal implant connections after a fatigue test.

Authors:  Paulo Cesar Pinheiro Feitosa; Ana Paula Barbosa de Lima; Laís Regiane Silva-Concílio; William Cunha Brandt; Ana Christina Claro Neves
Journal:  Eur J Dent       Date:  2013-07

5.  The effect of repeated torque tightening on total lengths of implant abutments in different internal implant‒abutment connections.

Authors:  Fariba Saleh Saber; Nader Abolfazli; Soheil Jannatii Ataei; Mahsa Taghizade Motlagh; Vahede Gharekhani
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2017-06-21

6.  Influence of abutment design on retention of metal copings cemented to implants.

Authors:  Albano Porto da Cunha; Glauco Pereira Moysés; Ana Christina Claro Neves; Rafael Pino Vitti; Flávia Cardoso da Rosa Goulart; Laís Regiane da Siva-Concílio
Journal:  Acta Biomater Odontol Scand       Date:  2016-01-26

7.  Effect of different angulations and collar lengths of conical hybrid implant abutment on screw loosening after dynamic cyclic loading.

Authors:  Mai Ahmed Yousry El-Sheikh; Tamer Mohamed Nasr Mostafa; Mohamed Maamoun El-Sheikh
Journal:  Int J Implant Dent       Date:  2018-12-03

8.  Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test.

Authors:  Caio Marques Martins; Elimário Venturin Ramos; Simone Kreve; Geraldo Alberto Pinheiro de Carvalho; Aline Batista Gonçalves Franco; Luís Guilherme Scavone de Macedo; Alecsandro de Moura Silva; Sérgio Candido Dias
Journal:  Dent Res J (Isfahan)       Date:  2019 Mar-Apr

9.  Evaluation of reverse torque values and failure loads of three different abutment designs with internal connections.

Authors:  Saied Nokar; Habib Hajimiragha; Leyla Sadighpour; Azam Sadat Mostafavi
Journal:  Dent Res J (Isfahan)       Date:  2020-12-10
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.