Literature DB >> 11853067

Cytokine removal with a large pore cellulose triacetate filter: an ex vivo study.

S Uchino1, R Bellomo, D Goldsmith, P Davenport, L Cole, I Baldwin, S Panagiotopoulos, F Tipping, C Ronco, P Everard.   

Abstract

OBJECTIVE: To test the hypothesis that hemofiltration using a new large pore cellulose triacetate hemofilter can achieve effective ultrafiltration of cytokines.
DESIGN: Ex-vivo study.
SETTING: Laboratory of Intensive Care Unit in tertiary hospital.
SUBJECTS: Six healthy volunteers.
INTERVENTIONS: Blood from 6 volunteers was incubated for 4 hours with 1 mg of endotoxin and then circulated through a closed hemofiltration circuit with a large pore cellulose triacetate hemofilter (nominal cut-off point: 60 kilodaltons). Hemofiltration was conducted at 1 L/h or 6 L/h of ultrafiltrate (UF) flow at the start of extra-corporeal circulation, and after 2 and 4 hours. Samples were taken from the arterial, venous and UF sampling ports.
MEASUREMENTS AND MAIN RESULTS: IL-Ibeta, IL-6, IL-8, IL-10, TNFalpha, and albumin were measured. Sieving coefficients (SC) above 0.6 were achieved for IL-Ibeta and IL-6 and SCs above 0.3 were achieved for IL-8 and TNF-alpha at 1 L/h. Sieving coefficients of all cytokines (except IL-10, p=0.22) were reduced when the ultrafiltration rate was increased from IL/h to 6 L/h (p<0.01), but the increase in ultrafiltration rate resulted in an overall increase in the clearance of all cytokines (p<0.001). The highest SC for albumin was 0.07 at 4 hours at 1 L/h, and fell to 0.01 at 6 L/h. The SCs for IL-8 fell at 4 hours (p<0.01), but the SCs for other cytokines did not change. No adsorption of cytokines and albumin was observed.
CONCLUSION: High volume hemofiltration (HVHF) using a new large pore cellulose triacetate filter achieved cytokine clearances greater than those reported with currently available hemo filters.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11853067     DOI: 10.1177/039139880202500105

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  3 in total

Review 1.  Hemoperfusion: technical aspects and state of the art.

Authors:  Claudio Ronco; Rinaldo Bellomo
Journal:  Crit Care       Date:  2022-05-12       Impact factor: 19.334

2.  Source and Purity of Dengue-Viral Preparations Impact Requirement for Enhancing Antibody to Induce Elevated IL-1β Secretion: A Primary Human Monocyte Model.

Authors:  Justin B Callaway; Scott A Smith; Douglas G Widman; Karen P McKinnon; Frank Scholle; Gregory D Sempowski; Dirk P Dittmer; James E Crowe; Aravinda M de Silva; Jenny P-Y Ting
Journal:  PLoS One       Date:  2015-08-24       Impact factor: 3.240

3.  Treatment of Acute Kidney Injury Complicating Septic Shock with EMiC2 High-cutoff Hemofilter: Case Series.

Authors:  Hoi-Ping Shum; King-Chung Chan; Wing-Wa Yan; Tak Mao Chan
Journal:  Indian J Crit Care Med       Date:  2017-11
  3 in total