The wear of metal-on-metal total hip prostheses measured in a hip simulator.

S C Scholes1, S M Green, A Unsworth.   

Abstract

New generation metal-on-metal prostheses have been introduced to try and overcome the problem of osteolysis often attributed to the wear particles of the polyethylene component of conventional metal-on-ultra-high molecular weight polyethylene (UHMWPE) joints. The wear rates of four metal-on-metal joints (two different clearances) were assessed along with that of a conventional metal-on-UHMWPE joint. Friction measurements of the metal-on-metal joints were taken before and after the wear test and compared. Two distinct wear phases were discernible for all the metal-on-metal joints: an initial wear phase up to 0.5 x 10(6) cycles and then a lower steady state wear phase. The steady state wear rate of the 22 microm radial clearance metal-on-metal joint was lower than that for the 40 microm radial clearance joint, although this difference was not found to be significant (p > 0.15). The wear rates for all the joints tested were consistent with other simulator studies. The friction factors produced by each joint were found to decrease significantly after wear testing (p < 0.05).

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11848384     DOI: 10.1243/0954411011536118

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  3 in total

1.  What are the risks accompanying the reduced wear benefit of low-clearance hip resurfacing?

Authors:  Joseph Daniel; Hena Ziaee; Amir Kamali; Chandra Pradhan; Derek McMinn
Journal:  Clin Orthop Relat Res       Date:  2012-07-24       Impact factor: 4.176

2.  Does diametrical clearance influence the wear of Pinnacle hip implants?

Authors:  Sean Bergiers; Harry Hothi; Johann Henckel; Antti Eskelinen; John Skinner; Alister Hart
Journal:  Bone Joint Res       Date:  2020-08-25       Impact factor: 5.853

3.  Wear of the Charité® lumbar intervertebral disc replacement investigated using an electro-mechanical spine simulator.

Authors:  Parshia Moghadas; Aziza Mahomed; Duncan E T Shepherd; David W L Hukins
Journal:  Proc Inst Mech Eng H       Date:  2015-03       Impact factor: 1.617

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.