| Literature DB >> 11839887 |
S Serge Barold1, Stéphane Garrigue, Jacques Clémenty.
Abstract
This report describes two cases of far-field sensing of the P wave by the ventricular channel of conventional DDD pacemakers programmed to a relatively high sensitivity to promote sensing of ventricular extrasystoles. Ventricular pacing was maintained in both cases when the ventricular channel was not inhibited. One case was caused by displacement of a bipolar ventricular lead towards the right ventricular inflow tract. The other case occurred only in relation to an atrial extrasystole in the absence of ventricular lead displacement.Entities:
Mesh:
Year: 2002 PMID: 11839887 DOI: 10.1023/a:1014132408411
Source DB: PubMed Journal: J Interv Card Electrophysiol ISSN: 1383-875X Impact factor: 1.900