Literature DB >> 11839887

Far-field P-wave sensing by the right ventricular lead of conventional dual chamber pacemakers.

S Serge Barold1, Stéphane Garrigue, Jacques Clémenty.   

Abstract

This report describes two cases of far-field sensing of the P wave by the ventricular channel of conventional DDD pacemakers programmed to a relatively high sensitivity to promote sensing of ventricular extrasystoles. Ventricular pacing was maintained in both cases when the ventricular channel was not inhibited. One case was caused by displacement of a bipolar ventricular lead towards the right ventricular inflow tract. The other case occurred only in relation to an atrial extrasystole in the absence of ventricular lead displacement.

Entities:  

Mesh:

Year:  2002        PMID: 11839887     DOI: 10.1023/a:1014132408411

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  5 in total

Review 1.  Oversensing by single-chamber pacemakers: mechanisms, diagnosis, and treatment.

Authors:  S S Barold; M D Falkoff; L S Ong; R A Heinle
Journal:  Cardiol Clin       Date:  1985-11       Impact factor: 2.213

2.  P-wave sensing in VVI pacemakers: useful or a problem?

Authors:  L M van Gelder; M I el Gamal; C H Tielen
Journal:  Pacing Clin Electrophysiol       Date:  1988-10       Impact factor: 1.976

3.  Inhibition of transvenous pacing through the coronary sinus by the atrial P wave: diagnosis with the aid of isoproterenol.

Authors:  M Weinstock; R DeGuia; M Daniell; J I Haft
Journal:  Chest       Date:  1971-05       Impact factor: 9.410

4.  P wave oversensing in a unipolar VVI pacemaker.

Authors:  B M van Gelder; F A Bracke; M I El Gamal
Journal:  Pacing Clin Electrophysiol       Date:  1995-02       Impact factor: 1.976

5.  Oversensing of pacemakers in the bipolar pacing configuration: paradoxic resolution with programming to unipolar sensing.

Authors:  U M Ben-Zur; J N Gross; M H Goldberger; P L Tilkemeier; A K Weyman; S Furman
Journal:  Am Heart J       Date:  1994-09       Impact factor: 4.749

  5 in total
  1 in total

1.  Inhibition of ventricular output in a dual chamber pacemaker with normal pacing and sensing parameters: What is the mechanism?

Authors:  Debabrata Bera; Radhey Shyam Joshi; Suchit Majumder; Sanjeev S Mukherjee
Journal:  Indian Pacing Electrophysiol J       Date:  2021-09-04
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.