| Literature DB >> 11839020 |
Abstract
Efforts are underway by our laboratory to develop a microcalorimeter that can be routinely used for radioactivity standardizations of nuclides that decay by pure beta-emission or by low-Z electron capture. The prototype calorimeter consists of a cryostat with two temperature-controlled stages and a base stage that are operated at nominal temperatures of 8 K. A unique aspect of the calorimeter's design is the ability to repeatedly engage and disengage the radioactive source from the second stage heat path using a magnetically activated elevator. The measurement of the total power from a radioactive sample is obtained from the difference in the second stage power with and without the source in place. As a result of extensive performance evaluations using 90Sr-90Y and 32P brachytherapy seeds as well as with an internal calibration heater, many initial design flaws have been identified and are being addressed.Entities:
Year: 2002 PMID: 11839020 DOI: 10.1016/s0969-8043(01)00192-0
Source DB: PubMed Journal: Appl Radiat Isot ISSN: 0969-8043 Impact factor: 1.513