| Literature DB >> 11836051 |
P Favetta1, C Allombert, C Breysse, C Dufresne, J Guitton, J Bureau.
Abstract
The stability of ceftazidime in 5% dextrose injection and 0.9% sodium chloride injection when stored in a different disposable infusion device was determined. Solutions of ceftazidime 40 mg/ml were used to fill the drug administration devices. Stability was determined for both 5% dextrose injection and 0.9% sodium chloride injection solutions at 37 degrees C in four disposable infusion devices. Ceftazidime and its mean degradation product, pyridine, were simultaneously assayed in triplicate by a stability-indicating high-performance liquid chromatographic (HPLC) method. This method was simple, sensitive (limit of quantitation (LOQ), 2 ng injected for both compounds), rapid (run time was 7 min) and precise (mean recovery was 100.5+/-2.9 and 103.6+/-1.9% for pyridine and ceftazidime, respectively). The ceftazidime stability in the 5% dextrose solution was lower than in the 0.9% sodium chloride solution. When stored at 37 degrees C in a disposable infusion device, the stability of the ceftazidime is included in large hourly range, depending strongly on the manufacturer. The stability of ceftazidime exceed 19 h in none studied cases. The pyridine formed in 24 h was in the range of 100-400 mg depending on devices and infusions.Entities:
Mesh:
Substances:
Year: 2002 PMID: 11836051 DOI: 10.1016/s0731-7085(01)00588-x
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935