Comparison of cardiac output and circulatory blood volumes by transpulmonary thermo-dye dilution and transcutaneous indocyanine green measurement in critically ill patients.

Samir G Sakka1, Konrad Reinhart, Karl Wegscheider, Andreas Meier-Hellmann.   

Abstract

OBJECTIVE: We prospectively studied the agreement between transpulmonary aortic fiberoptic-based and pulse dye densitometry (PDD) measurements of cardiac output and circulatory blood volumes.
DESIGN: Prospective clinical study.
SETTING: Operative ICU of a university hospital. PATIENTS: Sixteen critically ill, deeply sedated patients receiving mechanical ventilation with ARDS (n = 8), sepsis/septic shock (n = 6), subarachnoid hemorrhage (n = 1), and severe head injury (n = 1). MEASUREMENTS AND
RESULTS: Each patient received a 4F aortic catheter with an integrated fiberoptic and thermistor that was connected to a computer system for automatic calculation of the transpulmonary indicator dilution (TPID) technique for the measurement of cardiac output (COTPID), intrathoracic blood volume (ITBV), and total blood volume measured by TPID technique (TBVTPID). In each patient, an indocyanine green sensor was attached to one nasal wing and connected to an analyzer for the PDD measurement of cardiac output (COPDD), central blood volume (CBV), and TBV measured by PDD (TBVPDD). For all first measurements, linear regression analysis between COTPID and COPDD revealed that COPDD = 0.63 x COTPID + 3.69 (L/min) [r = 0.64, p = 0.008]. Mean bias between both techniques was - 0.8 L/min (SD, 1.7 L/min). Correlations between ITBV/CBV (r = 0.52) and TBVTPID/TBVPDD were only moderate: TBVPDD = 0.74 x TBVTPID + 2,362 (mL) [r = 0.60, p = 0.015; mean bias, - 999 mL; SD, 1,353 mL]. Over all 55 measurements, TPID measurements were on average 11.5% (cardiac output) and 17.6% (TBV) higher than PDD measurements. The differences between both measurements ranged from - 58 to 81% (cardiac output) and from - 47 to 82% (TBV; 95% reference ranges). The main source of variation were the intraindividual differences, resulting in different peaks and trends in the patients time courses depending on which measurement method was used.
CONCLUSION: PDD measurement of cardiac output and circulatory blood volumes agrees moderately with transpulmonary thermo-dye dilution technique in critically ill patients.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11834672     DOI: 10.1378/chest.121.2.559

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  22 in total

1.  Closed-Loop Intravenous Drug Administration Using Photoplethysmography.

Authors:  George W Carpenter; Holly G Myers; Eric A Sherer; Katie A Evans; D Patrick O'Neal
Journal:  IEEE J Transl Eng Health Med       Date:  2018-11-01       Impact factor: 3.316

2.  Photodeposition of silver can result in metal-enhanced fluorescence.

Authors:  Chris D Geddes; Alexandr Parfenov; Joseph R Lakowicz
Journal:  Appl Spectrosc       Date:  2003-05       Impact factor: 2.388

3.  Electrochemical and Laser Deposition of Silver for Use in Metal-Enhanced Fluorescence.

Authors:  Chris D Geddes; Alexandr Parfenov; David Roll; Jiyu Fang; Joseph R Lakowicz
Journal:  Langmuir       Date:  2003-07-22       Impact factor: 3.882

4.  Neurotoxic effects of indocyanine green -cerebellar granule cell culture viability study.

Authors:  Beata Toczylowska; Elzbieta Zieminska; Grazyna Goch; Daniel Milej; Anna Gerega; Adam Liebert
Journal:  Biomed Opt Express       Date:  2014-02-19       Impact factor: 3.732

5.  Cardiac output measurement by pulse dye densitometry: comparison with pulmonary artery thermodilution in post-cardiac surgery patients.

Authors:  Martijn Kroon; A B Johan Groeneveld; Yvo M Smulders
Journal:  J Clin Monit Comput       Date:  2006-01-25       Impact factor: 2.502

6.  Stability of cirrhotic systemic hemodynamics ensures sufficient splanchnic blood flow after living-donor liver transplantation in adult recipients with liver cirrhosis.

Authors:  Tomohide Hori; Shintaro Yagi; Taku Iida; Kentaro Taniguchi; Kentaro Yamagiwa; Chiduru Yamamoto; Takashi Hasegawa; Koichiro Yamakado; Takuma Kato; Kanako Saito; Linan Wang; Mie Torii; Yukinobu Hori; Kan Takeda; Kazuo Maruyama; Shinji Uemoto
Journal:  World J Gastroenterol       Date:  2007-11-28       Impact factor: 5.742

7.  Indocyanine green alters transepithelial electrical parameters of the distal colon.

Authors:  Burhan Hameed; David M Smith; Jon J Verrechio; J David Schmidt; Leesa E Gillooley; Mary Carmen Valenzano; Simon A Lewis; James M Mullin
Journal:  Dig Dis Sci       Date:  2004-09       Impact factor: 3.199

Review 8.  Systemic hemodynamics in advanced cirrhosis: Concerns during perioperative period of liver transplantation.

Authors:  Tomohide Hori; Yasuhiro Ogura; Yasuharu Onishi; Hideya Kamei; Nobuhiko Kurata; Motoshi Kainuma; Hideo Takahashi; Shogo Suzuki; Takashi Ichikawa; Shoko Mizuno; Tadashi Aoyama; Yuki Ishida; Takahiro Hirai; Tomoko Hayashi; Kazuko Hasegawa; Hiromu Takeichi; Atsunobu Ota; Yasuhiro Kodera; Hiroyuki Sugimoto; Taku Iida; Shintaro Yagi; Kentaro Taniguchi; Shinji Uemoto
Journal:  World J Hepatol       Date:  2016-09-08

9.  Roughened silver electrodes for use in metal-enhanced fluorescence.

Authors:  Chris D Geddes; Alexandr Parfenov; David Roll; Ignacy Gryczynski; Joanna Malicka; Joseph R Lakowicz
Journal:  Spectrochim Acta A Mol Biomol Spectrosc       Date:  2004-07       Impact factor: 4.098

10.  Blood volume measurement with indocyanine green pulse spectrophotometry: dose and site of dye administration.

Authors:  Menno R Germans; Philip C de Witt Hamer; Leonard J van Boven; Koos A H Zwinderman; Gerrit J Bouma
Journal:  Acta Neurochir (Wien)       Date:  2009-09-04       Impact factor: 2.216

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.