Literature DB >> 11830627

Use of the percutaneous thrombolytic device for the treatment of thrombosed pseudoaneurysms during mechanical thrombectomy of hemodialysis grafts.

Anthony N Hein1, Thomas M Vesely.   

Abstract

This article describes a method to remove thrombus from pseudoaneurysms associated with hemodialysis grafts during a percutaneous mechanical thrombectomy procedure. This effective and easily performed technique involves manual compression of the pseudoaneurysm during activation of the Arrow Percutaneous Thrombolytic Device. The authors describe their clinical experience with use of this technique to treat 18 patients during percutaneous mechanical thrombectomy procedures. In addition, the development, morphology, and management of hemodialysis graft-related pseudoaneurysms are discussed.

Entities:  

Mesh:

Year:  2002        PMID: 11830627     DOI: 10.1016/s1051-0443(07)61939-3

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  2 in total

1.  Treatment of hemodialysis graft pseudoaneurysms with stent grafts: institutional experience and review of the literature.

Authors:  Lauren R Pandolfe; Angelo P Malamis; Kenneth Pierce; Marc A Borge
Journal:  Semin Intervent Radiol       Date:  2009-06       Impact factor: 1.513

2.  Occlusion of a rapidly expanding hemodialysis graft pseudoaneurysm with placement of a stent graft.

Authors:  Arie Moszkowicz; George Behrens; Sebouh Gueyikian; Nilesh H Patel; Hector Ferral
Journal:  Semin Intervent Radiol       Date:  2007-03       Impact factor: 1.513

  2 in total