[Interference of hydroxocobalamine treatment on commun biochemical determinations].

P Vest1, C Renaudeau, S Tellal, C Ragot, C Renard.   

Abstract

Hydroxocobalamin, antidote of the cyanide poisoning, is a red and water soluble pigment, which colors biological mediums. The interferences caused by this molecule are studied for plasma pools overloaded with hydroxocobalamin for concentrations from 0 to 1,125 mg.L-1. The results of 16 biochemical parameters on Vitros 750 XRC (Ortho-clinical Diagnostics) and of 23 biochemical parameters on Cobas Integra 700 (Roche Diagnostics) are examined. With this analytical system, unconjugated and conjugated bilirubin, creatinin, and CK results are changed as soon as plasma concentration of hydroxocobalamin reachs 124 mg.L-1. For the Vitros 750 XRC, ASAT, iron, creatinin, phosphorus are more sensitive to this interference. But, other results are changed for the two analytical systems for superior hydroxocobalamin concentrations.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11830394

Source DB:  PubMed          Journal:  Ann Biol Clin (Paris)        ISSN: 0003-3898            Impact factor:   0.459