A direct comparison of amperometric gas sensors with gas-diffusion and ion-exchange membrane based electrodes.

René Knake1, Patrick Jacquinot, Peter C Hauser.   

Abstract

The effect of the nature of the working electrode used in amperometric gas sensors on the performance criteria of sensitivity, detection limit, gas flow rate and humidity dependence was evaluated. The arrangement based on metallized ion-exchange membranes (Nafion) was compared with gas-diffusion electrodes based on porous poly(tetrafluoroethylene) (PTFE) with metallic electrodes deposited on the rear side. Two representative analyte gases were chosen: SO2, which has fast reaction kinetics, and NO, which has slow reaction kinetics. It was found that both types of electrodes showed a similar performance. A dependence on the flow rate of the sample gas was found in both cases. The sensitivities were higher for the ion-exchange membrane-backed electrodes; however, the 3sigma detection limits were all in the lower ppb range and for NO were significantly lower on the Nafion membrane than on the PTFE membrane. The Nafion electrode was found to show a dependence on the relative humidity of the gas stream, but not the PTFE-based electrode.

Entities:  

Year:  2002        PMID: 11827376     DOI: 10.1039/b105527p

Source DB:  PubMed          Journal:  Analyst        ISSN: 0003-2654            Impact factor:   4.616


  1 in total

1.  Methane-oxygen electrochemical coupling in an ionic liquid: a robust sensor for simultaneous quantification.

Authors:  Zhe Wang; Min Guo; Gary A Baker; Joseph R Stetter; Lu Lin; Andrew J Mason; Xiangqun Zeng
Journal:  Analyst       Date:  2014-08-05       Impact factor: 4.616

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.