| Literature DB >> 11824076 |
Abstract
Extensive blood loss requires adequate volume replacement. However the infused volume cannot be adequately warmed especially when high infusion rates are necessary. Subsequently, hypothermia develops and results in hemodynamic instability and coagulopathy. The Rapid Infusion System (RIS) allows high infusion rates (up to 1.5 l/min) while at the same time guaranteeing sufficient warming. The efficacy of the RIS was investigated in 43 consecutive patients who required a massive transfusion. The average volume transfused in these patients was 31.7 +/- 4.5 l (minimum: 7.8 l; maximum: 165.3 l) which is equal to an average exchange of 6.4 times the circulating blood volume (maximum: 39.4 blood volumes). The replacement of such high blood volumes has not yet been published in a series of patients. Despite these high transfusion rates, the body core temperature was maintained at 35.85 +/- 0.1 degrees C. Only five patients had a body core temperature below 34 degrees C, all were trauma patients and four of these five patients already had a preoperative temperature below 34 degrees C. The mortality in this study was 28%, which is markedly reduced in comparison to previous publications although they all considered at patients with significantly less blood loss. Maintaining normothermia and normovolemia by the use of the RIS may explain the improved outcome.Entities:
Mesh:
Year: 2001 PMID: 11824076 DOI: 10.1007/s00101-001-0244-y
Source DB: PubMed Journal: Anaesthesist ISSN: 0003-2417 Impact factor: 1.041