| Literature DB >> 118179 |
A L Barry, D Amsterdam, M B Coyle, E H Gerlach, C Thornsberry, R W Hawkinson.
Abstract
A collaborative study was undertaken to evaluate a simple, convenient device which expedites inoculum standardization for antimicrobial disk susceptibility tests. The Inocupac system (Minnesota Mining & Manufacturing [3M] Co., St. Paul, Minn.) was used to perform disk tests in parallel with the standard Bauer-Kirby method. Five investigators tested 100 selected isolates, each in triplicate. Inter- and intralaboratory precision of both disk procedures was essentially comparable. The Inocupac system failed to consistently provide satisfactory growth with some streptococci, but when testing common gram-negative bacilli, staphylococci, and most enterococci, the Inocupac system gave zones about the same size (+/-2 mm) as the Bauer-Kirby procedure. Interpretive agreement between the two test systems varied from 90 to 99% with different antimicrobial agents, and repeated tests with the Bauer-Kirby method demonstrated the same degree of interpretive agreement. The Inocupac system is a valid alternative method for inoculating disk susceptibility tests.Entities:
Mesh:
Substances:
Year: 1979 PMID: 118179 PMCID: PMC273292 DOI: 10.1128/jcm.10.6.910-918.1979
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948