Literature DB >> 11817874

Excessive heat generation during cutting of cement in the Robodoc hip-revision procedure.

M Nogler1, M Krismer, C Haid, M Ogon, C Bach, C Wimmer.   

Abstract

The ROBODOC system is a promising new method for removing cement with high-speed milling. Heat is generated during the milling process. This study was designed to measure temperatures in the cutting area, and to assess the risk of heat injury and the effectiveness of irrigation. We measured temperatures at the bone-cement cutting area in three experimental settings, two involving the proximal area comprising a cement mantle, and one the distal cement plug beneath the prosthesis. Without cooling facilities, a mean temperature of 94 degrees C was measured in proximal areas. However, this could effectively be reduced below 70 degrees C with irrigation. In the area of the distal cement plug, we measured a mean temperature of 172 degrees C without irrigation. In this area, the integrated irrigation system with an additional high-flow irrigation system could not guarantee cooling to an acceptable temperature of below 70 degrees C since the irrigation stream was impeded by the cutter in the narrow cavity. We need an integrated irrigation device that guarantees continuous cooling at the cutting interface in front of the cutter.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11817874     DOI: 10.1080/000164701317269012

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  5 in total

1.  New aspects and approaches in pre-operative planning of hip reconstruction: a computer simulation.

Authors:  Marco Viceconti; Alessandro Chiarini; Debora Testi; Fulvia Taddei; Barbara Bordini; Francesco Traina; Aldo Toni
Journal:  Langenbecks Arch Surg       Date:  2004-01-15       Impact factor: 3.445

2.  The Evolution of Computer-Assisted Total Hip Arthroplasty and Relevant Applications.

Authors:  Jun-Dong Chang; In-Sung Kim; Atul M Bhardwaj; Ramachandra N Badami
Journal:  Hip Pelvis       Date:  2017-03-06

3.  Aerosol morphology and particle size distribution in orthopaedic bone machining: a laboratory worst-case contamination simulation. Is high-speed bone machining potentially harmful by pollution and quality schemes and what measures could be taken for prevention?

Authors:  David Putzer; Dietmar Dammerer; Cora Huber; Harald Boschert; Martin Thaler; Michael Nogler
Journal:  Int Orthop       Date:  2022-04-18       Impact factor: 3.479

Review 4.  Computer-assisted orthopaedic surgery and robotic surgery in total hip arthroplasty.

Authors:  Nobuhiko Sugano
Journal:  Clin Orthop Surg       Date:  2013-02-20

5.  Cement Removal from the Femur Using the ROBODOC System in Revision Total Hip Arthroplasty.

Authors:  Mitsuyoshi Yamamura; Nobuo Nakamura; Hidenobu Miki; Takashi Nishii; Nobuhiko Sugano
Journal:  Adv Orthop       Date:  2013-10-22
  5 in total