| Literature DB >> 11816646 |
Abstract
This report describes an asymptomatic patient with a DDDR pacemaker system (Inos Cls, Biotronik) that showed intermittent mode reversion to VOO and sometimes to VVI without a detectable source of electromagnetic interference and oversensing. Patients with this Biotronik model should be carefully assessed along with a surface ECG before placing the programming head.Entities:
Mesh:
Year: 2001 PMID: 11816646 DOI: 10.1046/j.1460-9592.2001.01704.x
Source DB: PubMed Journal: Pacing Clin Electrophysiol ISSN: 0147-8389 Impact factor: 1.976