Intermittent mode reversion to VOO of a dual chamber pacemaker model.

O Erdogan1, A Altun, G Ozbay.   

Abstract

This report describes an asymptomatic patient with a DDDR pacemaker system (Inos Cls, Biotronik) that showed intermittent mode reversion to VOO and sometimes to VVI without a detectable source of electromagnetic interference and oversensing. Patients with this Biotronik model should be carefully assessed along with a surface ECG before placing the programming head.