Literature DB >> 11805009

Inaccuracy of lipid measurements with the portable Cholestech L.D.X analyzer in patients with hypercholesterolemia.

James H Stein1, Cynthia M Carlsson, Kristi Papcke-Benson, Jean A Einerson, Patrick E McBride, Donald A Wiebe.   

Abstract

BACKGROUND: Although total cholesterol concentrations measured by portable lipid analyzers have acceptable bias and precision in young and middle-aged adults, clinically relevant differences in HDL-cholesterol (HDL-C) and triglyceride values have been described. Furthermore, the accuracy of portable lipid analyzers in older hyperlipidemic individuals, who have a high incidence of coronary heart disease, has not been validated. This study determined the biases and variability in portable lipid measurements in older patients with hypercholesterolemia and related them to National Cholesterol Education Program Adult Treatment Panel III guidelines.
METHODS: Participants were > or =70 years of age with fasting serum LDL-cholesterol (LDL-C) concentrations > 1.40 g/L. Fasting fingerstick samples were analyzed on a Cholestech L.D.X desktop analyzer. Antecubital venous samples were analyzed in a proficiency-certified clinical laboratory.
RESULTS: Portable measurements systematically overestimated triglycerides (0.296 g/L; P <0.001) and HDL-C (0.015 g/L; P = 0.026). LDL-C concentrations were underestimated (0.043 g/L; P = 0.046). Total and non-HDL cholesterol calculations based on the portable lipid device provided unbiased estimates, but wide variability was present. Significant variability in lipid determinations limited their clinical usefulness in individual patients, especially because 2 SD of the mean bias between the laboratory and the portable determinations of LDL-C and non-HDL cholesterol exceeded the 0.30 g/L cutoff that defines treatment targets in the current lipid guidelines.
CONCLUSIONS: Lipid values obtained from portable lipid analyzers may be useful for screening, but they should not be used to make clinical decisions regarding the diagnosis and management of dyslipidemia in individual patients.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11805009

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  6 in total

Review 1.  Specialized lipid profiles.

Authors:  Karen E Friday
Journal:  Curr Atheroscler Rep       Date:  2002-09       Impact factor: 5.113

2.  Point-of-care testing for the analysis of lipid panels: primary care diagnostic technology update.

Authors:  Annette Plüddemann; Matthew Thompson; Christopher P Price; Jane Wolstenholme; Carl Heneghan
Journal:  Br J Gen Pract       Date:  2012-03       Impact factor: 5.386

Review 3.  Screening tests for adults with intellectual disabilities.

Authors:  Joanne E Wilkinson; Larry Culpepper; Mary Cerreto
Journal:  J Am Board Fam Med       Date:  2007 Jul-Aug       Impact factor: 2.657

4.  Clinical utility of a fingerstick technology to identify individuals with abnormal blood lipids and high-sensitivity C-reactive protein levels.

Authors:  Parin Parikh; Heidi Mochari; Lori Mosca
Journal:  Am J Health Promot       Date:  2009 Mar-Apr

5.  Effect of dietary marine microalgae (schizochytrium) powder on egg production, blood lipid profiles, egg quality, and Fatty Acid composition of egg yolk in layers.

Authors:  J H Park; S D Upadhaya; I H Kim
Journal:  Asian-Australas J Anim Sci       Date:  2015-03       Impact factor: 2.509

6.  Establishing Waist-to-Height Ratio Standards from Criterion-Referenced BMI Using ROC Curves in Low-Income Children.

Authors:  Ryan D Burns; Timothy A Brusseau; Yi Fang; You Fu; James C Hannon
Journal:  J Obes       Date:  2016-11-03
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.