European concerted action on anticoagulation. Use of plasma samples to derive international sensitivity index for whole-blood prothrombin time monitors.

Leon Poller1, Michelle Keown, Nikhil Chauhan, Anton M H P van den Besselaar, Joyce Meeuwisse-Braun, Armando Tripodi, Mariagrazia Clerici, Jørgen Jespersen.   

Abstract

BACKGROUND: To simplify International Sensitivity Index (ISI) calibration, the possibility of substituting fresh plasma for fresh whole-blood samples with point-of-care testing (POCT) whole-blood monitors was investigated in a three-center study of three different POCT systems.
METHODS: A modified full WHO calibration procedure based on 20 healthy controls and 60 coumarin-treated patients was performed on three monitoring systems with whole-blood and plasma samples against plasma tested using the European Concerted Action on Anticoagulation (ECAA) rabbit reference plain thromboplastin and the manual prothrombin time (PT) method.
RESULTS: With one of the three systems, the mean ISI was 1.51 for whole blood and 1.49 for plasma; with the second system, the mean ISI was 1.08 for both whole blood and plasma. With the third system, however, the difference between the mean ISI for whole blood and that for plasma was greater (1.15 and 1.01, respectively). Overall, the precision of the calibrations was less than with traditional manual plasma PT testing.
CONCLUSIONS: Provided that an appropriate calcium chloride concentration is used, the plasma PT results can be used for accurate ISI calibration of two of these three whole-blood POCT systems. Precision criteria need to be modified for POCT monitors.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11805005

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  4 in total

1.  European Concerted Action on Anticoagulation (ECAA). An assessment of lyophilised plasmas for ISI calibration of CoaguChek and TAS whole blood prothrombin time monitors.

Authors:  L Poller; M Keown; N Chauhan; A M H P van den Besselaar; A Tripodi; C Shiach; J Jespersen
Journal:  J Clin Pathol       Date:  2003-02       Impact factor: 3.411

2.  Reliability of international normalised ratios from two point of care test systems: comparison with conventional methods.

Authors:  Leon Poller; Michelle Keown; Nikhil Chauhan; Anton M H P van den Besselaar; Armando Tripodi; Caroline Shiach; Jorgen Jespersen
Journal:  BMJ       Date:  2003-07-05

3.  Fluorochrome-linked immunoassay for functional analysis of the mannose binding lectin complement pathway to the level of C3 cleavage.

Authors:  Mary C Walsh; Lisa A Shaffer; Benjamin J Guikema; Simon C Body; Stanton K Shernan; Amanda A Fox; Charles D Collard; Michael Fung; Ronald P Taylor; Gregory L Stahl
Journal:  J Immunol Methods       Date:  2007-05-11       Impact factor: 2.303

4.  Monitoring of anticoagulant therapy in heart disease: considerations for the current assays.

Authors:  Mohammadali Boroumand; Hamidreza Goodarzynejad
Journal:  J Tehran Heart Cent       Date:  2010-05-31