Validation of the scanning -gamma-ray telescope for in vivo dosimetry and boron measurements during BNCT.

W F Verbakel1.   

Abstract

Gamma-ray telescope scans of a box phantom with inhomogeneous boron concentrations have proven the feasibility of in vivo measurements of different boron distributions in the head of a patient during boron neutron capture therapy (BNCT). Small structures with enhanced boron concentration can be reconstructed in a head phantom, even if the brain compartment of the phantom is surrounded by a skin layer with a ten times higher boron concentration. The motor-controlled telescope can scan the head/phantom, detecting boron and hydrogen prompt y-rays emitted at neutron capture reactions with a two-dimensional spatial resolution of 14 mm full width at half maximum. For reconstruction of the boron concentrations from the measured y-ray detection rates, a mathematical reconstruction algorithm is derived and discussed. Proper reconstruction requires position-dependent y-ray measurements combined with treatment planning programme calculations of the thermal neutron distribution. In a head phantom, in which the brain and the skull (bulk) were represented using a homogeneous boron distribution of 5.2 +/- 0.5 ppm 10B, surrounded by a skin layer with a ten times higher boron concentration, the bulk concentration was reconstructed to 4.7 +/- 0.3 ppm 10B. Telescope scans along and perpendicular to the beam axis showed the influence of inhomogeneities with a high boron concentration such as skin and a simulated blood vessel, respectively with a low boron concentration such as white matter. The profiles of the boron and hydrogen y-ray detection rates indicate how future patient measurements can be interpreted. In clinical trials, the telescope can then be used to investigate the averaged boron concentration in the bulk of a patient and local enhanced boron concentrations (e.g. in tumour tissue) in order to relate the measured boron dose distributions to the clinical effects of BNCT. Simultaneously, it can serve as quality control of the dosimetry during the irradiation.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11768505     DOI: 10.1088/0031-9155/46/12/314

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

Review 1.  Computational dosimetry and treatment planning considerations for neutron capture therapy.

Authors:  David W Nigg
Journal:  J Neurooncol       Date:  2003 Mar-Apr       Impact factor: 4.130

Review 2.  Clinical Viability of Boron Neutron Capture Therapy for Personalized Radiation Treatment.

Authors:  Dominika Skwierawska; José Antonio López-Valverde; Marcin Balcerzyk; Antonio Leal
Journal:  Cancers (Basel)       Date:  2022-06-10       Impact factor: 6.575

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.