Microabrasion--a simple method to assess surface degradation of UHMWPE following sterilisation and ageing.

F J Buchanan1, P H Shipway.   

Abstract

The ageing behaviour of ultra-high molecular weight polyethylene (UHMWPE) has been studied following gamma irradiation (25 or 40 kGy) in air. Accelerated ageing procedures used elevated temperature (70 degrees C) and/or pressurised oxygen (5bar). Shelf-aged UHMWPE was also studied. The variation in surface density and mechanical properties were determined following the various sterilisation and ageing treatments. Microabrasive wear testing was also performed. Wear rates were found to correlate well with stress at break for sterilised and aged UHMWPE but not with elongation to failure. It is proposed that the wear mechanism is fracture dominated and occurs following some disentanglement of the polymer chains. Wear also depends upon embrittlement of the surface layer due to its processing and ageing. Elongation to failure in a tensile test is not a good measure of this embrittlement whereas the microabrasion test provides more surface sensitive information concerning this property.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11762860     DOI: 10.1016/s0142-9612(01)00083-7

Source DB:  PubMed          Journal:  Biomaterials        ISSN: 0142-9612            Impact factor:   12.479


  2 in total

1.  Chemical stability of oil-infused polyethylene.

Authors:  Fedra P Zaribaf; Harinderjit S Gill; Elise C Pegg
Journal:  J Biomater Appl       Date:  2020-12-26       Impact factor: 2.646

2.  Damages of the tibial post in constrained total knee prostheses in the early postoperative course - a scanning electron microscopic study of polyethylene inlays.

Authors:  Adrian Skwara; Carsten O Tibesku; Rudolf Reichelt; Susanne Fuchs-Winkelmann
Journal:  BMC Musculoskelet Disord       Date:  2008-06-11       Impact factor: 2.362

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.