Literature DB >> 11761411

Clinical experience with a dual-chamber implantable cardioverter defibrillator to treat atrial tachyarrhythmias.

M R Gold1, N Sulke, D S Schwartzman, R Mehra, D E Euler.   

Abstract

INTRODUCTION: This study evaluated the safety and efficacy of a new dual-chamber implantable cardioverter defibrillator (ICD) to detect and treat atrial tachyarrhythmias in patients with drug-refractory atrial fibrillation (AF) and no indication for a ventricular ICD. METHODS AND
RESULTS: A dual-chamber ICD (Medtronic 7250 Jewel AF) was implanted in 144 of 146 patients. The device discriminates atrial tachycardia from AF based on cycle length and regularity, and uses atrial overdrive pacing as well as shocks to terminate tachyarrhythmia episodes. Patients were followed for an average of 12.6+/-6.2 months. Use of antiarrhythmic drugs was 63% at baseline and did not change over time. Kaplan-Meier estimates of 12-month complication-free survival, device therapy survival, and patient survival were 85%, 91%, and 98%, respectively. Positive predictive accuracy of spontaneous atrial tachyarrhythmia detection was 99%, while atrial overdrive pacing and shocks terminated 40% and 87% of treated episodes, respectively. Median duration of successfully treated episodes was 8.9 minutes versus 144 minutes for the therapy failures. There was no reduction in the use of patient-activated shock therapy over time; at the 12-month follow-up evaluation, 94% of patients were in sinus rhythm. Ventricular tachyarrhythmias (67 episodes) were detected and appropriately treated in 7.6% of patients.
CONCLUSION: This dual-chamber ICD appears to be safe and well tolerated in patients with drug-refractory symptomatic atrial tachyarrhythmias. The device, used in combination with drugs, effectively treats atrial tachyarrhythmias with pacing and/or shock therapies and decreases the median episode duration. In addition, the device protects from ventricular tachyarrhythmias in patients with AF and structural heart disease.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11761411     DOI: 10.1046/j.1540-8167.2001.01247.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  15 in total

Review 1.  Management of atrial fibrillation in bradyarrhythmias.

Authors:  Giuseppe Boriani; Luigi Padeletti
Journal:  Nat Rev Cardiol       Date:  2015-03-17       Impact factor: 32.419

2.  Device-based therapies for atrial fibrillation.

Authors:  Gregory K Bruce; Paul A Friedman
Journal:  Curr Treat Options Cardiovasc Med       Date:  2005-10

3.  Is dual defibrillator better than conventional DDD pacing in brady-tachy syndrome? Results of the ICARUS Trial (Internal Cardioversion Antitachypacing and Prevention: Resource Utilization Study).

Authors:  Massimo Santini; Renato Ricci; Carlo Pignalberi; Maurizio Russo; Barbara Magris; Nicoletta Grovale; Tiziana De Santo
Journal:  J Interv Card Electrophysiol       Date:  2006-01-18       Impact factor: 1.900

4.  Recommendations for pacemaker implantation for the treatment of atrial tachyarrhythmias and resynchronisation therapy for heart failure: A report from the task force on pacemaker indications of the Dutch Working Group on Cardiac Pacing.

Authors:  N M van Hemel; B Dijkman; W G de Voogt; W P Beukema; H A Bosker; C C de Cock; L J L M Jordaens; I C van Gelder; L M van Gelder; R van Mechelen; J H Ruiter; M I Sedney; L C Slegers
Journal:  Neth Heart J       Date:  2004-01       Impact factor: 2.380

5.  Implantable dual-chamber defibrillator for the selective treatment of spontaneous atrial and ventricular arrhythmias: arrhythmia incidence and device performance.

Authors:  Andreas Schuchert; Giuseppe Boriani; Christian Wollmann; Mauro Biffi; Martin Kühl; Johannes Sperzel; Sascha Stiller; Gianni Gasparini; Dirk Böcker
Journal:  J Interv Card Electrophysiol       Date:  2005-03       Impact factor: 1.900

Review 6.  Alternatives to warfarin for thromboembolism prophylaxis in nonrheumatic atrial fibrillation.

Authors:  Ramin Artang; Humberto Vidaillet
Journal:  J Interv Card Electrophysiol       Date:  2004       Impact factor: 1.900

7.  The circadian variation of atrial defibrillation thresholds.

Authors:  Andrew Robert John Mitchell; Eduardo Warman; Katie Schaaf; Neil Sulke
Journal:  J Interv Card Electrophysiol       Date:  2004-02       Impact factor: 1.900

8.  Dual defibrillator improves quality of life and decreases hospitalizations in patients with drug refractory atrial fibrillation.

Authors:  Renato Ricci; Aurelio Quesada; Carlo Pignalberi; Josè Roda; Marcello Disertori; Alessandro Capucci; Antonio Raviele; Massimo Santini
Journal:  J Interv Card Electrophysiol       Date:  2004-02       Impact factor: 1.900

Review 9.  Atrial fibrillation: the role of atrial defibrillation.

Authors:  Massimo Santini; Renato Ricci
Journal:  J Interv Card Electrophysiol       Date:  2003-10       Impact factor: 1.900

10.  The impact of atrial prevention and termination therapies on atrial tachyarrhythmia burden in patients receiving a dual-chamber defibrillator for ventricular arrhythmias.

Authors:  Paul A Friedman; John H Ip; Mohammad Jazayeri; John H Lobban; Brian Olshansky; Duo Zhou; Tracy Knowles; David E Euler
Journal:  J Interv Card Electrophysiol       Date:  2004-04       Impact factor: 1.900

View more