Literature DB >> 11754545

Potentiometric flow injection determination of glycerol in distilled spirits.

Ana Paula G Gervasio1, Eduardo P Borges, Elias A G Zagatto, Boaventura F Reis, Rui A S Lapa, José L F C Lima.   

Abstract

A single-line flow injection system including a tubular periodate-selective electrode without inner reference solution is proposed for glycerol determination in distilled spirits, based on oxidation of this polyol by periodate. Interferences due to 5.0 mg L(-1) Cu, 5000 mg L(-1) sucrose, and 3000 mg L(-1) fructose plus glucose were investigated. The procedure is characterized by a linear response for 20-500 mg L(-1) glycerol (r > 0.9999, n = 7), a relative standard deviation of results of <0.03, and an analytical throughput of 30 determinations per hour. Accuracy was assessed by applying the procedure to distilled spirits of sugarcane and grape already analyzed by HPLC; in addition, recoveries within 96 and 120% were obtained.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11754545     DOI: 10.1021/jf001417r

Source DB:  PubMed          Journal:  J Agric Food Chem        ISSN: 0021-8561            Impact factor:   5.279


  1 in total

1.  Identification of volatile compounds, quantification of glycerol and trace elements in distilled spirits produced in Mozambique.

Authors:  Manuel Carlos Minez Tábua; Wilder Douglas Santiago; Maísa Lamounier Magalhães; Vanuzia Rodrigues Fernandes Ferreira; Rafaela Magalhães Brandão; Maria Luisa Teixeira; Márcio Pozzobon Pedroso; Ana Maria de Resende Machado; David Lee Nelson; Maria das Graças Cardoso
Journal:  J Food Sci Technol       Date:  2019-09-10       Impact factor: 2.701