Literature DB >> 11754460

Impact of navigator timing on free-breathing submillimeter 3D coronary magnetic resonance angiography.

Elmar Spuentrup1, Warren J Manning, René M Botnar, Kraig V Kissinger, Matthias Stuber.   

Abstract

The purpose of this study was to investigate the impact of navigator timing on image quality in navigator-gated and real-time motion-corrected, free-breathing, three-dimensional (3D) coronary MR angiography (MRA) with submillimeter spatial image resolution. Both phantom and in vivo investigations were performed. 3D coronary MRA with real-time navigator technology was applied using variable navigator time delays (time delay between the navigator and imaging sequences) and varying spatial resolutions. Quantitative objective and subjective image quality parameters were assessed. For high-resolution imaging, reduced image quality was found as a function of increasing navigator time delay. Lower spatial resolution coronary MRA showed only minor sensitivity to navigator timing. These findings were consistent among volunteers and phantom experiments. In conclusion, for submillimeter navigator-gated and real-time motion-corrected 3D coronary MRA, shortening the time delay between the navigator and the imaging portion of the sequence becomes increasingly important for improved spatial resolution. Copyright 2002 Wiley-Liss, Inc.

Mesh:

Year:  2002        PMID: 11754460     DOI: 10.1002/mrm.10026

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  15 in total

1.  Free-breathing cardiac MR stress perfusion with real-time slice tracking.

Authors:  Tamer A Basha; Sébastien Roujol; Kraig V Kissinger; Beth Goddu; Sophie Berg; Warren J Manning; Reza Nezafat
Journal:  Magn Reson Med       Date:  2013-10-07       Impact factor: 4.668

Review 2.  Coronary magnetic resonance imaging: visualization of the vessel lumen and the vessel wall and molecular imaging of arteriothrombosis.

Authors:  Elmar Spuentrup; Rene M Botnar
Journal:  Eur Radiol       Date:  2005-08-25       Impact factor: 5.315

3.  Pulmonary vein inflow artifact reduction for free-breathing left atrium late gadolinium enhancement.

Authors:  Mehdi H Moghari; Dana C Peters; Jouke Smink; Lois Goepfert; Kraig V Kissinger; Beth Goddu; Thomas H Hauser; Mark E Josephson; Warren J Manning; Reza Nezafat
Journal:  Magn Reson Med       Date:  2011-02-28       Impact factor: 4.668

4.  Three-dimensional heart locator and compressed sensing for whole-heart MR angiography.

Authors:  Mehdi H Moghari; David Annese; Tal Geva; Andrew J Powell
Journal:  Magn Reson Med       Date:  2015-06-10       Impact factor: 4.668

5.  Improved respiratory self-navigation for 3D radial acquisitions through the use of a pencil-beam 2D-T2 -prep for free-breathing, whole-heart coronary MRA.

Authors:  Andrew J Coristine; Jerome Chaptinel; Giulia Ginami; Gabriele Bonanno; Simone Coppo; Ruud B van Heeswijk; Davide Piccini; Matthias Stuber
Journal:  Magn Reson Med       Date:  2017-05-31       Impact factor: 4.668

6.  Improved border sharpness of post-infarct scar by a novel self-navigated free-breathing high-resolution 3D whole-heart inversion recovery magnetic resonance approach.

Authors:  Tobias Rutz; Davide Piccini; Simone Coppo; Jerome Chaptinel; Giulia Ginami; Gabriella Vincenti; Matthias Stuber; Juerg Schwitter
Journal:  Int J Cardiovasc Imaging       Date:  2016-08-22       Impact factor: 2.357

7.  A dual-projection respiratory self-gating technique for whole-heart coronary MRA.

Authors:  Peng Lai; Andrew C Larson; Xiaoming Bi; Renate Jerecic; Debiao Li
Journal:  J Magn Reson Imaging       Date:  2008-09       Impact factor: 4.813

8.  Free-breathing 3-dimensional steady-state free precession coronary magnetic resonance angiography: comparison of four navigator gating techniques.

Authors:  Thanh D Nguyen; Pascal Spincemaille; Matthew D Cham; Jonathan W Weinsaft; Martin R Prince; Yi Wang
Journal:  Magn Reson Imaging       Date:  2009-01-17       Impact factor: 2.546

9.  Prospective heart tracking for whole-heart magnetic resonance angiography.

Authors:  Mehdi H Moghari; Tal Geva; Andrew J Powell
Journal:  Magn Reson Med       Date:  2016-02-04       Impact factor: 4.668

10.  Free-breathing 3D cardiac MRI using iterative image-based respiratory motion correction.

Authors:  Mehdi H Moghari; Sébastien Roujol; Raymond H Chan; Susie N Hong; Natalie Bello; Markus Henningsson; Long H Ngo; Beth Goddu; Lois Goepfert; Kraig V Kissinger; Warren J Manning; Reza Nezafat
Journal:  Magn Reson Med       Date:  2012-11-06       Impact factor: 4.668