Rapid liquid chromatography-tandem mass spectrometry method for routine analysis of cyclosporin A over an extended concentration range.

Brian G Keevil1, David P Tierney, Donald P Cooper, Michael R Morris.   

Abstract

BACKGROUND: Cyclosporin A (CsA) is commonly measured by immunoassay techniques that have a limited analytical range. The consequence of this is that low CsA concentrations that may be clinically significant are difficult to measure and that high concentrations require sample dilution, which introduces error and increases cost. More specific assays, such as HPLC, do not have the required turnaround times for busy transplant clinics.
METHODS: CsA was measured in whole blood from 180 cardiac and lung transplant recipients by a liquid chromatography--tandem mass spectrometry (MS) assay, and the results were compared with the Dade Behring Emit assay. Proteins were precipitated with acetonitrile containing ascomycin as internal standard. We used isocratic elution on a Supelco CN column (33 x 3.0 mm; 3- microm bead size) with a mobile phase of 65% aqueous acetonitrile containing ammonium acetate (2 mmol/L) and formic acid (1 g/L), at a flow rate of 0.5 mL/min, with a sample injection volume of 6 microL. We used positive-ion electrospray MS to monitor the ammonium adducts of the compounds of interest decomposing under controlled conditions to the most dominant fragments of the individual molecules. Calibration curves used linear least-squares regression with 1/x weighting.
RESULTS: Maximum sensitivity was obtained by monitoring fragmentation of the ammonium adducts m/z 1220-->m/z 1203 for CsA and m/z 809-->m/z 765 for ascomycin. Sample throughput, including preparation time, was 30 samples in 1.5 h with an injection-to-injection cycle time of 1.5 min. The calibration curve was linear to 5000 microg/L, with a detection limit of 0.03 microg/L and a limit of quantification of 1 microg/L. Regression analysis [tandem MS method (y) and Emit assay (x)] yielded a slope of 1.09 (+/- 0.03), an intercept of 6.2 (+/- 4.5) microg/L, and S(y/x) = 27 microg/L.
CONCLUSIONS: Tandem MS assay is a realistic alternative to immunoassay for the routine monitoring of CsA in transplant recipients. Its wide dynamic range has utility for pharmacokinetic studies of CsA.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11751540

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  2 in total

1.  A phase I trial of cyclosporine for hospitalized patients with COVID-19.

Authors:  Emily A Blumberg; Julia Han Noll; Pablo Tebas; Joseph A Fraietta; Ian Frank; Amy Marshall; Anne Chew; Elizabeth A Veloso; Alison Carulli; Walter Rogal; Avery L Gaymon; Aliza H Schmidt; Tiffany Barnette; Renee Jurek; Rene Martins; Briana M Hudson; Kalyan Chavda; Christina M Bailey; Sarah E Church; Hooman Noorchashm; Wei-Ting Hwang; Carl H June; Elizabeth O Hexner
Journal:  JCI Insight       Date:  2022-06-08

2.  Could Salivary Cyclosporine Dosage Replace the Whole Blood Cyclosporine Measurements in Renal Transplant Patients?

Authors:  Samia Aboujaoude; Marwan Masri; Sylvana Rizk; Souleima Chamat; Jad Farha; Zeina Majzoub
Journal:  J Int Soc Prev Community Dent       Date:  2017-05-22
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.