The flow rate of the elastomeric balloon infusor is influenced by the internal pressure of the infusor.

I S Chung1, H S Cho, J A Kim, K H Lee.   

Abstract

The purpose of this study was to evaluate the flow rate of elastomeric balloon infusor composed of rubber or silicone materials. Two models were studied: the Baxter Twoday Infusor and Advance Silicone Infusor. Each infusion device has a preset flow rate of 2 mL/hr. Fifteen units of each device were filled with 100 mL of normal saline. The flow rate was measured gravimetically using an electronic balance. The internal pressure of the balloon infusor was measured with a Digital pressure meter via the pressure transducer. The monitored internal pressure of the two kinds of infusor was not maintained uniformly during the entire delivery period, which was divided into 3 phases: the phase of decreasing, maintained, and increasing pressure during the delivery of drug. Both devices initially infused at a relatively high rate, followed by a somewhat steady flow rate. The flow rate distinctly increased in the small residual volume. The flow rate of the balloon infusor used in this study was not sustained uniformly during the entire delivery period and was in proportion to the internal pressure of the infusor regardless of the materials.

Entities:  

Mesh:

Year:  2001        PMID: 11748348      PMCID: PMC3054806          DOI: 10.3346/jkms.2001.16.6.702

Source DB:  PubMed          Journal:  J Korean Med Sci        ISSN: 1011-8934            Impact factor:   2.153


  4 in total

1.  The efficacy of elastomeric patient-control module when connected to a balloon pump for postoperative epidural analgesia: A randomized, noninferiority trial.

Authors:  Myung Hwa Kim; Yon Hee Shim; Min-Soo Kim; Yang-Sik Shin; Hyun Joo Lee; Jeong Soo Lee
Journal:  Medicine (Baltimore)       Date:  2017-01       Impact factor: 1.889

Review 2.  Newer drug delivery systems in anesthesia.

Authors:  Sona Dave; Deepa Shriyan; Pinakin Gujjar
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2017 Apr-Jun

3.  The flow Rate Accuracy of Elastomeric Infusion Pumps After Repeated Filling.

Authors:  Masood Mohseni; Amin Ebneshahidi
Journal:  Anesth Pain Med       Date:  2014-04-07

Review 4.  Stability of Antimicrobials in Elastomeric Pumps: A Systematic Review.

Authors:  Beatriz Fernández-Rubio; Paula Del Valle-Moreno; Laura Herrera-Hidalgo; Alicia Gutiérrez-Valencia; Rafael Luque-Márquez; Luis E López-Cortés; José María Gutiérrez-Urbón; Sonia Luque-Pardos; Aurora Fernández-Polo; María V Gil-Navarro
Journal:  Antibiotics (Basel)       Date:  2021-12-30
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.