| Literature DB >> 11747193 |
N Al-Mubarak1, G S Roubin, J J Vitek, S S Iyer, G New, M B Leon.
Abstract
A 61-year-old man underwent carotid stenting with the distal-balloon protection system for symptomatic carotid artery stenosis. During the procedure, progressive elevation of the systolic blood pressure occurred, reaching a peak of 220 mm Hg immediately following deflation of the distal balloon. This was associated with severe headaches and progressive deterioration in the mental status to a coma. Head CT scan showed massive subarachnoidal hemorrhage contralateral to the stented side and a secondary intracerebral hemorrhage. Despite immediate successful blood pressure control, his condition deteriorated and he died 2 days later. Copyright 2001 Wiley-Liss, Inc.Entities:
Mesh:
Year: 2001 PMID: 11747193 DOI: 10.1002/ccd.1324