| Literature DB >> 11747009 |
V M Mai1, A A Bankier, P V Prasad, W Li, P Storey, R R Edelman, Q Chen.
Abstract
Magnetic resonance ventilation-perfusion (V/Q) imaging has been demonstrated using oxygen and arterial spin labeling techniques. Inhaled oxygen is used as a paramagnetic contrast agent in ventilation imaging using a multiple inversion recovery (MIR) approach. Pulmonary perfusion imaging is conducted using a flow-sensitive alternating inversion recovery with an extra radiofrequency pulse (FAIRER) technique. A half Fourier single-short turbo spin echo (HASTE) sequence is used for data acquisition in both techniques. V/Q imaging was performed in ten of the twenty volunteers, while either ventilation or perfusion was imaged in the other ten. This V/Q imaging scheme is completely noninvasive, does not involve ionized radiation, and shows promising potential for clinical use in the diagnosis of lung diseases such as pulmonary embolism. Copyright 2001 Wiley-Liss, Inc.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11747009 DOI: 10.1002/jmri.1221
Source DB: PubMed Journal: J Magn Reson Imaging ISSN: 1053-1807 Impact factor: 4.813