| Literature DB >> 11742733 |
S Myler1, S D Collyer, K A Bridge, S P J Higson.
Abstract
An outer ultra-thin-polydimethyldichlorosiloxane film composite membrane has been used as the outer covering barrier in an amperometric glucose oxidase enzyme electrode biosensor. The composite membrane was formed via the condensation polymerisation of dimethyldichlorosilane at the surface of a host porous alumina membrane. Homogeneous polydimethyldichlorosiloxane films of <100 nm thickness acted as effective substrate diffusional barriers and were supported by the underlying porous alumina surface. Glucose and oxygen permeability coefficients were determined using diffusion chamber apparatus. Polysiloxane composite membranes were found to offer some screening functionality towards anionic biological interferents such as ascorbate. On exposure to blood an approximate 25% signal drift was observed during the first 2 h exposure to blood; after this time responses remained almost stable. Whole blood glucose determinations showed a close correlation (r(2)=0.98) to analyses performed via standard hospital analyses.Entities:
Mesh:
Substances:
Year: 2002 PMID: 11742733 DOI: 10.1016/s0956-5663(01)00265-2
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618