Cardiac pacemakers. Does radiation therapy affect performance?

B J Walz, R E Reder, J O Pastore, P Littman, R Johnson.   

Abstract

Representative deman cardiac pacemakers were tested with several common radiation therapy devices and ancillary equipment. The experimental set-up essentially duplicated the dosimetric and electroconductive environment of the radiation therapy patient bearing a pacemaker. Observations of pulse width and interval confirmed that pacemaker output was generally not appreciably affected by the simulated treatments. Occasionally, changes were caused by ancillary equipment and start-up of treatment. We conclude that although interference hazards are sometimes seen, radiation therapy is not contraindicated in patients wearing pacemakers.

Entities:  

Mesh:

Year:  1975        PMID: 1174228     DOI: 10.1001/jama.234.1.72

Source DB:  PubMed          Journal:  JAMA        ISSN: 0098-7484            Impact factor:   56.272


  1 in total