Assessment of the use of the laryngeal tube for cardiopulmonary resuscitation in a manikin.

H V Genzwuerker1, T Finteis, D Slabschi, J Groeschel, K Ellinger.   

Abstract

During 60 3-min CPR sequences, the face mask, laryngeal tube and tracheal tube were compared using an Ambu Megacode Trainer. Ten 3-min sequences each were performed for both a combination of the face mask and laryngeal tube with a bag-valve device (compression-ventilation ratio 5:1). With continuous chest compressions, ten 3-min CPR sequences each were performed for a combination of the laryngeal tube and tracheal tube with a bag-valve device and ten 3-min CPR sequences each for a combination of the laryngeal tube and tracheal tube with an automatic transport ventilator. Signs of gastric inflation occurred only with the face mask. Ventilation with the laryngeal tube was significantly better than with the face mask and comparable to the tracheal tube during ventilation with the bag-valve device and with the automatic transport ventilator. Chest compressions caused a significant decrease in tidal volumes during ventilation with the automatic transport ventilator. The findings of this study support the idea of the laryngeal tube as a new adjunct for emergency airway management, but will have to be verified during clinical practice.

Entities:  

Mesh:

Year:  2001        PMID: 11738781     DOI: 10.1016/s0300-9572(01)00410-5

Source DB:  PubMed          Journal:  Resuscitation        ISSN: 0300-9572            Impact factor:   5.262


  4 in total

1.  Effect of the laryngeal tube on the no-flow-time in a simulated two rescuer basic life support setting with inexperienced users.

Authors:  J Schröder; M Bucher; O Meyer
Journal:  Med Klin Intensivmed Notfmed       Date:  2015-09-15       Impact factor: 0.840

2.  Using a laryngeal tube during cardiac arrest reduces "no flow time" in a manikin study: a comparison between laryngeal tube and endotracheal tube.

Authors:  Christoph H R Wiese; Utz Bartels; Anna Bergmann; Ingo Bergmann; Jan Bahr; Bernhard M Graf
Journal:  Wien Klin Wochenschr       Date:  2008       Impact factor: 1.704

3.  [First clinical experiences with the new LTS. A laryngeal tube with an oesophageal drain].

Authors:  H Genzwürker; T Finteis; J Hinkelbein; K Ellinger
Journal:  Anaesthesist       Date:  2003-06-19       Impact factor: 1.041

4.  Laryngeal tube use in out-of-hospital cardiac arrest by paramedics in Norway.

Authors:  Geir A Sunde; Guttorm Brattebø; Terje Odegården; Dag F Kjernlie; Emma Rødne; Jon-Kenneth Heltne
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2012-12-18       Impact factor: 2.953

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.