IgG antibodies typical for extrinsic allergic alveolitis--an inter-laboratory quality assessment.

W Aberer1, M Woltsche, I Woltsche-Kahr, B Kränke.   

Abstract

BACKGROUND: Determination of specific IgG antibodies is important for the diagnosis of extrinsic allergic alveolitis (EAA). Various evaluations have however shown, that current methodology lacks sufficient standardization in that the employment of different sources of extracts and techniques makes a comparison of data from one laboratory to another almost impossible.
OBJECTIVE: The aim of this study is to establish an external quality control system and to analyse, what the explanations for the different outcomes from various laboratories might be.
METHODS: In the past 4 years 5 sera from patients suffering from EAA or healthy controls were sent every 6 months to 11 different allergy laboratories in Austria. The determination of specific IgG antibodies against antigens that are typical for this disease were requested. Results were gained with the method routinely used in the respective laboratory, and then sent back to the reference center for statistical evaluation. Precipitating techniques were used in 8 laboratories during the first mailings, but were gradually exchanged by automated ELISA systems being employed in 8 laboratories in the last mailing.
RESULTS: 1127 values were determined in 105 expectedly positive sera and 1003 in 94 negative samples. Of the 562 values obtained with precipitation techniques in positive sera, only 52.0% were reported to be positive, and the results varied considerably among laboratories and antigens. In contrast, 93.3% were positive with commercially available ELISA techniques, with 92.3% for the EnzyDex System and even 95.5% for the UniCAP System. Regarding the specificity however, 93.0% of the expected negative results were correct negative using precipitation methods, whereas merely 75.2% were negative with the EnzyDex System and only 22.5% using the UniCAP System. Moreover 35.8% of the results using this latter method were false-positive.
CONCLUSIONS: The traditional precipitation techniques proved not only technically difficult to perform, but also unreliable, difficult to reproduce, insensitive and impractical in daily laboratory work. They suffer from that many draw backs, that their use in daily routine cannot be recommended any more. Automated ELISA systems seem to fulfill the criteria for a routine technique concerning handling, automation, and quality criteria like sensitivity quite well, but not for specificity. Both techniques urgently need external standardization in order to make the results comparable among the different systems and methods; the danger of potentially false-positive results, pretending sensitizations that might be clinically irrelevant in several cases, is high.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11726311

Source DB:  PubMed          Journal:  Eur J Med Res        ISSN: 0949-2321            Impact factor:   2.175


  2 in total

Review 1.  Clinical immunology review series: an approach to the use of the immunology laboratory in the diagnosis of clinical allergy.

Authors:  P Williams; W A C Sewell; C Bunn; R Pumphrey; G Read; S Jolles
Journal:  Clin Exp Immunol       Date:  2008-07       Impact factor: 4.330

Review 2.  Hypersensitivity pneumonitis.

Authors:  Yves Lacasse; Yvon Cormier
Journal:  Orphanet J Rare Dis       Date:  2006-07-03       Impact factor: 4.123

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.