First results using automated epifluorescence microscopy to detect Escherichia coli and Staphylococcus epidermidis in WBC-reduced platelet concentrates.

M Seaver1, J C Crookston, D C Roselle, S J Wagner.   

Abstract

BACKGROUND: Many methods have been tested for the detection of bacterial contamination in platelets. However, only those using molecular biology or cell culturing consistently detect contamination at levels below 10(5) bacteria per mL. This report describes the initial investigation into an alternative method that offers the possibilities of high sensitivity and rapid response while using available laboratory equipment and supplies. This method relies on a fluorescent nucleic acid stain, which preferentially stains bacteria but not platelets, and automated epifluorescence microscopy for rapid analysis. Measurements in WBC-reduced platelet concentrates (PCs) contaminated with bacteria are reported at concentrations between 10(3) and 10(6) bacteria per mL. STUDY DESIGN AND METHODS: Staphylococcus epidermidis or Escherichia coli was inoculated into aliquots of WBC-reduced PCs on Days 2 through 5 of storage. Bacterially inoculated and control PCs were stained, platelets and residual WBCs were lysed, and 200 microL of sample was filtered onto black polycarbonate filters. All preparations were done in triplicate. An automated epifluorescence microscope examined approximately 2 percent of the area of each filter and used image analysis to select the fluorescent particles that should be counted as bacteria.
RESULTS: Samples containing 3 to 5 x 10(3) bacteria per mL produced about three times as many fluorescent particles classified as bacteria as the controls. Lower concentrations of S. epidermidis were detected because of higher fluorescence intensity. Simultaneous preparation of six samples requires about 35 minutes. Analysis of each prepared sample takes 10 minutes, for a total preparation and analysis time of about 95 minutes for 6 samples.
CONCLUSION: Low concentrations (<5 x 10(3) bacteria/mL) of deliberately inoculated S. epidermidis or E. coli can be measured quickly in WBC-depleted PCs by using a fluorescent nucleic acid stain, differential lysis, and automated microscopy. Continued refinement of the method, studies employing other bacterial strains, and further validations of assay performance are warranted.

Entities:  

Mesh:

Year:  2001        PMID: 11724977     DOI: 10.1046/j.1537-2995.2001.41111351.x

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  2 in total

1.  Microscopic detection of viable Staphylococcus epidermidis in peri-implant tissue in experimental biomaterial-associated infection, identified by bromodeoxyuridine incorporation.

Authors:  C A N Broekhuizen; M Sta; C M J E Vandenbroucke-Grauls; S A J Zaat
Journal:  Infect Immun       Date:  2010-01-04       Impact factor: 3.441

Review 2.  Protecting the blood supply from emerging pathogens: the role of pathogen inactivation.

Authors:  Jean Pierre Allain; Celso Bianco; Morris A Blajchman; Mark E Brecher; Michael Busch; David Leiby; Lily Lin; Susan Stramer
Journal:  Transfus Med Rev       Date:  2005-04